Deviations Between Transpalpebral Tonometry Using TGDc-01 and Goldmann Applanation Tonometry Depending on the IOP Level

Overview
Specialty Ophthalmology
Date 2005 Sep 15
PMID 16158308
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Recent comparisons between transpalpebral tonometry using TGDc-01 and Goldmann applanation tonometry were performed in populations with IOPs between 10 and 20 mmHg. The purpose of this study was to evaluate device deviations depending on different IOP levels (range 5-40 mmHg).

Methods: A total of 68 eyes of 68 patients were included and assigned to four IOP levels according to an initial applanation tonometry assessment: level I, <10 mmHg (n=8); level II, 10-19 mmHg (n=20); level III, 20-29 mmHg (n=20); and level IV, > or =30 mmHg (n=20). Two independent and randomized observers performed three replicate measurements per eye-observer 1 using TGDc-01 tonometry, and observer 2 using Goldmann applanation tonometry. Intraindividual deviations between measurement results were investigated concerning clinical relevance by medians and quartiles, concerning statistical significance by pairwise sign tests; p values <0.05 indicate local statistical significance.

Results: In patients with initial IOP > or =20 mmHg, TGDc-01-based tonometry significantly underestimated the IOP as based on Goldmann applanation tonometry (p<0.001). This effect increased with increasing IOP: IOP level III median difference (TGDc-01 - Goldmann) -1.3 mmHg (interquartile range, -2.5, -0.4), IOP level IV median difference -2.7 mmHg (-3.7, -1.0). In patients with initial IOP <10 mmHg, an at least gradual underestimation by TGDc-01 tonometry (p=0.219; median difference, -0.6, -1.6, 0) was observed. A total 18% of patients showed device deviations > +/-3 mmHg, and even 35% of those patients with initial IOP > or =30 mmHg.

Conclusions: TGDc-01-based tonometry demonstrated an increasing underestimation of IOP with increasing IOP levels when compared with the current standard method of Goldmann applanation tonometry.

Citing Articles

Easyton® transpalpebral versus Perkins applanation tonometry in different populations.

Montolio-Marzo E, Morales-Fernandez L, Saenz-Frances San Baldomero F, Garcia-Saenz S, Garcia-Feijoo J, Pinero D Int Ophthalmol. 2023; 43(10):3491-3497.

PMID: 37286775 DOI: 10.1007/s10792-023-02754-7.


Comparison of Pneumatonometry and Transpalpebral Tonometry Measurements of Intraocular Pressure during Scleral Lens Wear.

Fogt J, Nau C, Schornack M, Shorter E, Nau A, Harthan J Optom Vis Sci. 2020; 97(9):711-719.

PMID: 32941336 PMC: 7547891. DOI: 10.1097/OPX.0000000000001574.


Ageing and ocular surface immunity.

Mashaghi A, Hong J, Chauhan S, Dana R Br J Ophthalmol. 2016; 101(1):1-5.

PMID: 27378485 PMC: 5583682. DOI: 10.1136/bjophthalmol-2015-307848.


Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry.

Li Y, Shi J, Duan X, Fan F Graefes Arch Clin Exp Ophthalmol. 2010; 248(12):1765-70.

PMID: 20495818 DOI: 10.1007/s00417-009-1243-y.


[Measuring intraocular pressure by different methods].

Lamparter J, Hoffmann E Ophthalmologe. 2009; 106(8):676-82.

PMID: 19652979 DOI: 10.1007/s00347-009-1971-8.


References
1.
Doughty M, Zaman M . Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. 2000; 44(5):367-408. DOI: 10.1016/s0039-6257(00)00110-7. View

2.
Bland J, Altman D . Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. View

3.
GOLDMANN H, Schmidt T . [Applanation tonometry]. Ophthalmologica. 1957; 134(4):221-42. DOI: 10.1159/000303213. View

4.
Amm M, Hedderich J . [Transpalpebral tonometry with a digital tonometer in healthy eyes and after penetrating keratoplasty]. Ophthalmologe. 2004; 102(1):70-6. DOI: 10.1007/s00347-004-1082-5. View

5.
Quigley H, Langham M . Comparative intraocular pressure measurements with the pneumatonograph and Goldmann tonometer. Am J Ophthalmol. 1975; 80(2):266-73. DOI: 10.1016/0002-9394(75)90144-0. View