Clinical Comparison of the Oculab Tono-Pen to the Goldmann Applanation Tonometer

Overview
Journal Ophthalmology
Publisher Elsevier
Specialty Ophthalmology
Date 1987 Dec 1
PMID 3431824
Citations 32
Authors
Affiliations
Soon will be listed here.
Abstract

The Oculab Tono-Pen tonometer was compared with the Goldmann applanation tonometer through studies on 197 eyes of 103 patients with glaucoma or ocular hypertension to evaluate the reliability of the Tono-Pen in determining intraocular pressure (IOP). Patients were randomized into four groups and all applanations were done by the same investigator. The Tono-Pen initially used (Tono-Pen #1) failed to calibrate after being used on 77 eyes. Tono-Pen #2 was then obtained and tested. Each Tono-Pen showed high correlation with the Goldmann readings (r greater than or equal to 0.86 in both eyes). The overall sensitivity of the Tono-Pen in detecting IOPs of 21 mmHg or higher (as measured by the Goldmann instrument) was 62.1% in right eyes, 72.4% in left eyes, with corresponding specificities of 92.6% in the right and 97.1% in the left. Given the substantial increase in sensitivity noted with Tono-Pen #2, the Tono-Pen appears adequate for screening programs where an IOP of 21 mmHg or above is considered abnormal; however, at higher IOPs (greater than or equal to 30 mmHg) the Tono-Pens tended to underestimate Goldmann IOPs. At low IOPs (less than or equal to 9 mmHg) the Tono-Pens tended to overestimate the IOPs. Perhaps with further modifications of the instrument, it could be used for clinical diagnosis and management of glaucoma patients.

Citing Articles

Risk for Ocular Hypertension With Intravitreal Dexamethasone Implants in Black and White Patients With Diabetic Macular Edema.

Leung E, Alsoudi A, Skrehot H, Burkhead D, Adcock B, Behl I J Vitreoretin Dis. 2025; 24741264241309685.

PMID: 39790838 PMC: 11707758. DOI: 10.1177/24741264241309685.


Comparison of intraocular pressure measurements with the tono-pen, goldmann applanation tonometer, and noncontact tonometer in nonglaucomatous pseudophakic children.

Oskan Yalcin S, Kaplan A Int Ophthalmol. 2024; 44(1):285.

PMID: 38935310 DOI: 10.1007/s10792-024-03210-w.


Development of a biodegradable polymer-based implant to release dual drugs for post-operative management of cataract surgery.

Subhash N, Nair S, Srinivas S, Theruveethi N, Bhandary S, Guru B Drug Deliv Transl Res. 2024; 15(2):508-522.

PMID: 38696092 PMC: 11683021. DOI: 10.1007/s13346-024-01604-y.


Real-world agreement of same-visit Tono-Pen vs Goldmann applanation intraocular pressure measurements using electronic health records.

Gui H, Zhang Y, Chang R, Wang S Heliyon. 2023; 9(8):e18703.

PMID: 37576221 PMC: 10412763. DOI: 10.1016/j.heliyon.2023.e18703.


Telehealth and Screening Strategies in the Diagnosis and Management of Glaucoma.

Wong S, Tsai J J Clin Med. 2021; 10(16).

PMID: 34441748 PMC: 8396962. DOI: 10.3390/jcm10163452.