Comparison of the Tono-Pen to the Goldmann Applanation Tonometer

Overview
Journal Arch Ophthalmol
Specialty Ophthalmology
Date 1988 Jun 1
PMID 3369998
Citations 28
Authors
Affiliations
Soon will be listed here.
Abstract

Oculab Tono-Pen tonometry was compared with Goldmann applanation tonometry in 142 eyes of 71 patients without corneal problems. Overall we found 63% of the Tono-Pen readings to be within +/- 2 mm Hg of the Goldmann applanation tonometry readings and 77% to be within +/- 3 mm Hg, while 4% were within at least +/- 7 mm Hg difference. The correlation coefficient between the readings of the two instruments was .84. The mean value of the paired differences was 0.81 +/- 3.09 mm Hg overall, -1.78 +/- 1.86 mm Hg in the 4 to 10 mm Hg interval, 0.07 +/- 2.97 mm Hg in the 11 to 20 mm Hg interval, 1.27 +/- 2.62 mm Hg in the 21 to 30 mm Hg interval, and 4.15 +/- 2.28 mm Hg in the 31 to 45 mm Hg interval. We conclude that the Tono-Pen measures intraocular pressure in a manner that corresponds well to the Goldmann tonometer in the 11 to 20 mm Hg interval, and fairly well in the 4 to 10 mm Hg and 21 to 30 mm Hg intervals. It lacks good correspondence in the 31 to 45 mm Hg interval.

Citing Articles

Real-world agreement of same-visit Tono-Pen vs Goldmann applanation intraocular pressure measurements using electronic health records.

Gui H, Zhang Y, Chang R, Wang S Heliyon. 2023; 9(8):e18703.

PMID: 37576221 PMC: 10412763. DOI: 10.1016/j.heliyon.2023.e18703.


Intraocular Pressure Changes during Hemodiafiltration with Two different Concentrations of Sodium in the Dialysate.

Lerma C, Saavedra-Fuentes N, Ledesma-Gil J, Calderon-Juarez M, Perez-Grovas H, Lopez-Gil S Biology (Basel). 2022; 11(1).

PMID: 35053010 PMC: 8773306. DOI: 10.3390/biology11010012.


How to Measure Intraocular Pressure: An Updated Review of Various Tonometers.

Brusini P, Salvetat M, Zeppieri M J Clin Med. 2021; 10(17).

PMID: 34501306 PMC: 8456330. DOI: 10.3390/jcm10173860.


Telehealth and Screening Strategies in the Diagnosis and Management of Glaucoma.

Wong S, Tsai J J Clin Med. 2021; 10(16).

PMID: 34441748 PMC: 8396962. DOI: 10.3390/jcm10163452.


A comparison of NCT, Goldman application tonometry values with and without fluorescein.

Erdogan H, Akingol Z, Cam O, Sencan S Clin Ophthalmol. 2018; 12:2183-2188.

PMID: 30464378 PMC: 6211304. DOI: 10.2147/OPTH.S177870.