» Articles » PMID: 16122383

Noise Correlation in PET, CT, SPECT and PET/CT Data Evaluated Using Autocorrelation Function: a Phantom Study on Data, Reconstructed Using FBP and OSEM

Overview
Journal BMC Med Imaging
Publisher Biomed Central
Specialty Radiology
Date 2005 Aug 27
PMID 16122383
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Positron Emission Tomography (PET), Computed Tomography (CT), PET/CT and Single Photon Emission Tomography (SPECT) are non-invasive imaging tools used for creating two dimensional (2D) cross section images of three dimensional (3D) objects. PET and SPECT have the potential of providing functional or biochemical information by measuring distribution and kinetics of radiolabelled molecules, whereas CT visualizes X-ray density in tissues in the body. PET/CT provides fused images representing both functional and anatomical information with better precision in localization than PET alone. Images generated by these types of techniques are generally noisy, thereby impairing the imaging potential and affecting the precision in quantitative values derived from the images. It is crucial to explore and understand the properties of noise in these imaging techniques. Here we used autocorrelation function (ACF) specifically to describe noise correlation and its non-isotropic behaviour in experimentally generated images of PET, CT, PET/CT and SPECT.

Methods: Experiments were performed using phantoms with different shapes. In PET and PET/CT studies, data were acquired in 2D acquisition mode and reconstructed by both analytical filter back projection (FBP) and iterative, ordered subsets expectation maximisation (OSEM) methods. In the PET/CT studies, different magnitudes of X-ray dose in the transmission were employed by using different mA settings for the X-ray tube. In the CT studies, data were acquired using different slice thickness with and without applied dose reduction function and the images were reconstructed by FBP. SPECT studies were performed in 2D, reconstructed using FBP and OSEM, using post 3D filtering. ACF images were generated from the primary images, and profiles across the ACF images were used to describe the noise correlation in different directions. The variance of noise across the images was visualised as images and with profiles across these images.

Results: The most important finding was that the pattern of noise correlation is rotation symmetric or isotropic, independent of object shape in PET and PET/CT images reconstructed using the iterative method. This is, however, not the case in FBP images when the shape of phantom is not circular. Also CT images reconstructed using FBP show the same non-isotropic pattern independent of slice thickness and utilization of care dose function. SPECT images show an isotropic correlation of the noise independent of object shape or applied reconstruction algorithm. Noise in PET/CT images was identical independent of the applied X-ray dose in the transmission part (CT), indicating that the noise from transmission with the applied doses does not propagate into the PET images showing that the noise from the emission part is dominant. The results indicate that in human studies it is possible to utilize a low dose in transmission part while maintaining the noise behaviour and the quality of the images.

Conclusion: The combined effect of noise correlation for asymmetric objects and a varying noise variance across the image field significantly complicates the interpretation of the images when statistical methods are used, such as with statistical estimates of precision in average values, use of statistical parametric mapping methods and principal component analysis. Hence it is recommended that iterative reconstruction methods are used for such applications. However, it is possible to calculate the noise analytically in images reconstructed by FBP, while it is not possible to do the same calculation in images reconstructed by iterative methods. Therefore for performing statistical methods of analysis which depend on knowing the noise, FBP would be preferred.

Citing Articles

Inverse shielding and mutual exclusion for PET-MR hybrid imaging concerning induced positronium hyperfine splits radiations.

Wang K, Huq M Sci Rep. 2023; 13(1):20522.

PMID: 37993545 PMC: 10665340. DOI: 10.1038/s41598-023-44303-3.


Positive rate and quantification of amyloid pathology with [F]florbetapir in the urban Chinese population.

He K, Li B, Huang L, Zhao J, Hua F, Wang T Eur Radiol. 2023; 34(5):3331-3341.

PMID: 37889270 DOI: 10.1007/s00330-023-10366-z.


Numerical investigation reveals challenges in measuring the contrast recovery coefficients in PET.

Bayerlein R, Spencer B, Abdelhafez Y, Cherry S, Badawi R, Omidvari N Phys Med Biol. 2023; 68(21).

PMID: 37802064 PMC: 10798005. DOI: 10.1088/1361-6560/ad00fa.


A proper protocol for routine F-FDG uEXPLORER total-body PET/CT scans.

Hu H, Huang Y, Sun H, Zhou K, Jiang L, Zhong J EJNMMI Phys. 2023; 10(1):51.

PMID: 37695324 PMC: 10495295. DOI: 10.1186/s40658-023-00573-4.


A multi-modality physical phantom for mimicking tumor heterogeneity patterns in PET/CT and PET/MRI.

Valladares A, Beyer T, Papp L, Salomon E, Rausch I Med Phys. 2022; 49(9):5819-5829.

PMID: 35838056 PMC: 9543355. DOI: 10.1002/mp.15853.


References
1.
Alpert N, Chesler D, Correia J, Ackerman R, Chang J, Finklestein S . Estimation of the local statistical noise in emission computed tomography. IEEE Trans Med Imaging. 1982; 1(2):142-6. DOI: 10.1109/TMI.1982.4307561. View

2.
Gies M, Kalender W, Wolf H, Suess C . Dose reduction in CT by anatomically adapted tube current modulation. I. Simulation studies. Med Phys. 1999; 26(11):2235-47. DOI: 10.1118/1.598779. View

3.
Beyer T, Townsend D, Blodgett T . Dual-modality PET/CT tomography for clinical oncology. Q J Nucl Med. 2002; 46(1):24-34. View

4.
Faulkner K, Moores B . Analysis of x-ray computed tomography images using the noise power spectrum and autocorrelation function. Phys Med Biol. 1984; 29(11):1343-52. DOI: 10.1088/0031-9155/29/11/003. View

5.
Razifar P, Lubberink M, Schneider H, Langstrom B, Bengtsson E, Bergstrom M . Non-isotropic noise correlation in PET data reconstructed by FBP but not by OSEM demonstrated using auto-correlation function. BMC Med Imaging. 2005; 5(1):3. PMC: 1142517. DOI: 10.1186/1471-2342-5-3. View