» Articles » PMID: 15630604

Behavior of Arthroscopic Irrigation Systems

Overview
Publisher Wiley
Date 2005 Jan 5
PMID 15630604
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

In the literature, no consensus exists about optimal irrigation of joints during arthroscopic operations. The goal of this paper is to study the behavior of irrigation systems resulting in the proposal of guidelines for optimal irrigation. To this end, optimal irrigation is defined as the steady state of irrigation of a joint in which a sufficient positive intra-articular pressure and a sufficient flow are maintained. A model of the complete irrigation system was created to schematically elucidate the behavior of pump systems. Additionally, clinical experiments were performed during arthroscopic knee operations in which the pressure at different locations and the irrigation flow were measured. The combination of model prediction and clinical results could well be used to derive guidelines, since the clinical results, which showed considerable variation, were used to verify the model, and the model could be used to explain the typical trends. The main findings are twofold the set pressure is always higher than the intra-articular pressure, and the scope-sheath combination has a significant influence on irrigation control, because of its large restriction. Based on the results, we advice to increase the set pressure during active suction, and to include the sheath-scope combination in the control loop.

Citing Articles

Analysis of risk factors for procedure-related hemorrhage in rotator cuff repair surgery under shoulder arthroscopy.

Guo D, Wang J Medicine (Baltimore). 2023; 102(39):e35512.

PMID: 37773784 PMC: 10545304. DOI: 10.1097/MD.0000000000035512.


Computational Metrics Can Provide Quantitative Values to Characterize Arthroscopic Field of View.

Barnes R, Golden M, Borland D, Heckert R, Richardson M, Creighton R Arthrosc Sports Med Rehabil. 2022; 4(2):e403-e409.

PMID: 35494292 PMC: 9042744. DOI: 10.1016/j.asmr.2021.10.017.


Cost-Effectiveness Analysis Based on Intelligent Electronic Medical Arthroscopy for the Treatment of Varus Knee Osteoarthritis.

Liu C, Wang Z, Liu J, Xu Y J Healthc Eng. 2021; 2021:5569872.

PMID: 34035884 PMC: 8121568. DOI: 10.1155/2021/5569872.


Cost Effective Technique of Shoulder Arthroscopy Without the Use of Epinephrine in Irrigation Solution.

Stetson W, Morgan S, Polinsky S, Chung B, Hung N Arthrosc Tech. 2021; 10(2):e411-e418.

PMID: 33680773 PMC: 7917405. DOI: 10.1016/j.eats.2020.10.025.


[The effect of different continuous saline irrigation volume under arthroscopy on early postoperative pain and swelling of the knee].

Wang C, Yang P, Zhang D, Zhang Y, Shen Y, Li H Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2021; 35(2):178-182.

PMID: 33624470 PMC: 8171692. DOI: 10.7507/1002-1892.202006134.


References
1.
Weisleder L . Fluid pump systems for arthroscopy: a comparison of pressure control versus pressure and flow control. Arthroscopy. 1995; 11(5):591-5. DOI: 10.1016/0749-8063(95)90137-x. View

2.
Oretorp N, Elmersson S . Arthroscopy and irrigation control. Arthroscopy. 1986; 2(1):46-50. DOI: 10.1016/s0749-8063(86)80010-x. View

3.
Dolk T, Augustini B . Three irrigation systems for motorized arthroscopic surgery: a comparative experimental and clinical study. Arthroscopy. 1989; 5(4):307-14. DOI: 10.1016/0749-8063(89)90147-3. View

4.
Bergstrom R, Gillquist J . The use of an infusion pump in arthroscopy. Arthroscopy. 1986; 2(1):41-5. DOI: 10.1016/s0749-8063(86)80009-3. View

5.
Ewing J, Noe D, Kitaoka H, Askew M . Intra-articular pressures during arthroscopic knee surgery. Arthroscopy. 1986; 2(4):264-9. DOI: 10.1016/s0749-8063(86)80083-4. View