» Articles » PMID: 15327306

Pulstrodes: Triple Pulse Control of Potentiometric Sensors

Overview
Journal J Am Chem Soc
Specialty Chemistry
Date 2004 Aug 26
PMID 15327306
Citations 23
Authors
Affiliations
Soon will be listed here.
Abstract

Ion-selective electrode membranes based on hydrophobic materials doped with chemically selective host molecules are an attractive sensing technology but normally suffer from a limited sensitivity, given by the Nernst equation, and a direct reliance on the reference electrode potential, which makes miniaturization difficult. These fundamental problems are addressed here by imposing a multipulse electrochemical excitation signal onto ion-selective membranes that lack ion-exchange properties. Current pulses are responsible for the generation of ion fluxes in the direction of the membrane, which give reproducible super-Nernstian response slopes that originate from depletion processes at the membrane surface. Membranes may also be measured at zero current after this pulse, giving super-Nernstian response regions at lower concentrations. Difference potentials obtained from subsequent pulses give about 10-fold higher sensitivities than predicted on the basis of the Nernst equation.

Citing Articles

Chronopotentiometric sensors for antimicrobial peptide-based biosensing of Staphylococcus aureus.

Zhao J, Ding J, Luan F, Qin W Mikrochim Acta. 2024; 191(6):356.

PMID: 38811412 DOI: 10.1007/s00604-024-06410-4.


Catalytic Biosensors Operating under Quasi-Equilibrium Conditions for Mitigating the Changes in Substrate Diffusion.

Muhs A, Bobrowski T, Lielpetere A, Schuhmann W Angew Chem Int Ed Engl. 2022; 61(52):e202211559.

PMID: 36253337 PMC: 10099152. DOI: 10.1002/anie.202211559.


Unlocking All-Solid Ion Selective Electrodes: Prospects in Crop Detection.

Zhai J, Luo B, Li A, Dong H, Jin X, Wang X Sensors (Basel). 2022; 22(15).

PMID: 35898054 PMC: 9331676. DOI: 10.3390/s22155541.


Anodic Stripping Voltammetry on a Carbon-based Ion-Selective Electrode.

Park S, Maier C, Koley D Electrochim Acta. 2021; 390.

PMID: 34483347 PMC: 8415442. DOI: 10.1016/j.electacta.2021.138855.


Integrated multi-ISE arrays with improved sensitivity, accuracy and precision.

Wang C, Yuan H, Duan Z, Xiao D Sci Rep. 2017; 7:44771.

PMID: 28303939 PMC: 5356001. DOI: 10.1038/srep44771.