A Simple Method to Assess Accuracy of Deep Brain Stimulation Electrode Placement: Pre-operative Stereotactic CT + Postoperative MR Image Fusion
Overview
Authors
Affiliations
Objective: To describe a method for the measurement of the accuracy of deep brain stimulation (DBS) electrode placement with the use of image fusion technologies.
Patients And Methods: Ten consecutive patients suffering from movement disorders underwent DBS electrode placement. Postoperative MR images were fused with the pre-operative stereotactic CT. The placement error in the anteroposterior, lateral and vertical planes was calculated.
Results: The anteroposterior mean error +/- SD was 0.61 +/- 0.22 mm (range 0.2-0.9 mm). The lateral mean error +/- SD was 0.65 +/- 0.27 mm (range 0.2-2.2 mm). The vertical mean error +/- SD was 0.82 +/- 0.31 mm (range 0.3-1.6 mm).
Conclusions: This technique provides a simple and precise method for the evaluation of the accuracy of DBS electrode placement.
Rusche T, Kaufmann J, Voges J Hum Brain Mapp. 2021; 42(18):5888-5910.
PMID: 34528323 PMC: 8596959. DOI: 10.1002/hbm.25657.
Park H, Lim Y, Song E, Lee J, Park K, Park K J Clin Med. 2020; 9(9).
PMID: 32967337 PMC: 7564882. DOI: 10.3390/jcm9093044.
Kremer N, Oterdoom D, van Laar P, Pina-Fuentes D, van Laar T, Drost G Neuromodulation. 2019; 22(4):472-477.
PMID: 30629330 PMC: 6618091. DOI: 10.1111/ner.12918.
Van der Loo L, Schijns O, Hoogland G, Colon A, Wagner G, Dings J Acta Neurochir (Wien). 2017; 159(9):1733-1746.
PMID: 28676892 PMC: 5557874. DOI: 10.1007/s00701-017-3242-9.
Post-operative imaging in deep brain stimulation: A controversial issue.
Saleh C, Dooms G, Berthold C, Hertel F Neuroradiol J. 2016; 29(4):244-9.
PMID: 27029393 PMC: 4978322. DOI: 10.1177/1971400916639960.