Three-dimensional Color Doppler Sonography in Carotid Artery Stenosis
Overview
Authors
Affiliations
Background And Purpose: Color Doppler sonography (CDS) suffers from such disadvantages as high interobserver variability and problems with data presentation and storage. We therefore performed a prospective trial to assess the role of three-dimensional CDS (3D-CDS) in the evaluation of internal carotid artery stenosis (ICAS).
Methods: We included 32 consecutive patients with sonographically verified ICAS (30-99%) and 16 asymptomatic volunteers. All underwent CDS and 3D-CDS, and 23 patients also underwent intraarterial angiography. The sensitivity, specificity, positive predictive value, and negative predictive value for the detection of high-grade ICAS were analyzed by three blinded readers, who also graded the percentage of ICAS.
Results: CDS revealed 14 patients with mild to moderate ICAS and 18 patients with high-grade ICAS. Mean sensitivity of 3D-CDS was 81.5%, mean specificity was 98.9%, mean positive predictive value was 97.9%, and mean negative predictive value was 89.9%. 3D-CDS correlated significantly with CDS (mean r=0.85; P<.001) and angiography (mean r=0.57; P=.01).
Conclusion: 3D-CDS findings correlate with those obtained by CDS and angiography; 3D-CDS also offers excellent interobserver correlation, positive predictive value, and specificity, approaching 100% for the detection of high-grade ICAS.
Guidelines for ultrasound examination in ophthalmology. Part III: Color Doppler ultrasonography.
Modrzejewska M J Ultrason. 2019; 19(77):128-136.
PMID: 31355585 PMC: 6750311. DOI: 10.15557/JoU.2019.0019.
Pelz J, Weinreich A, Karlas T, Saur D PLoS One. 2017; 12(1):e0167500.
PMID: 28045903 PMC: 5207436. DOI: 10.1371/journal.pone.0167500.
MR plaque imaging of the carotid artery.
Watanabe Y, Nagayama M Neuroradiology. 2010; 52(4):253-74.
PMID: 20155353 DOI: 10.1007/s00234-010-0663-z.