Evaluation of Freehand B-Mode and Power-Mode 3D Ultrasound for Visualisation and Grading of Internal Carotid Artery Stenosis
Overview
Affiliations
Background: Currently, colour-coded duplex sonography (2D-CDS) is clinical standard for detection and grading of internal carotid artery stenosis (ICAS). However, unlike angiographic imaging modalities, 2D-CDS assesses ICAS by its hemodynamic effects rather than luminal changes. Aim of this study was to evaluate freehand 3D ultrasound (3DUS) for direct visualisation and quantification of ICAS.
Methods: Thirty-seven patients with 43 ICAS were examined with 2D-CDS as reference standard and with freehand B-mode respectively power-mode 3DUS. Stenotic value of 3D reconstructed ICAS was calculated as distal diameter respectively distal cross-sectional area (CSA) reduction percentage and compared with 2D-CDS.
Results: There was a trend but no significant difference in successful 3D reconstruction of ICAS between B-mode and power mode (examiner 1 {Ex1} 81% versus 93%, examiner 2 {Ex2} 84% versus 88%). Inter-rater agreement was best for power-mode 3DUS and assessment of stenotic value as distal CSA reduction percentage (intraclass correlation coefficient {ICC} 0.90) followed by power-mode 3DUS and distal diameter reduction percentage (ICC 0.81). Inter-rater agreement was poor for B-mode 3DUS (ICC, distal CSA reduction 0.36, distal diameter reduction 0.51). Intra-rater agreement for power-mode 3DUS was good for both measuring methods (ICC, distal CSA reduction 0.88 {Ex1} and 0.78 {Ex2}; ICC, distal diameter reduction 0.83 {Ex1} and 0.76 {Ex2}). In comparison to 2D-CDS inter-method agreement was good and clearly better for power-mode 3DUS (ICC, distal diameter reduction percentage: Ex1 0.85, Ex2 0.78; distal CSA reduction percentage: Ex1 0.63, Ex2 0.57) than for B-mode 3DUS (ICC, distal diameter reduction percentage: Ex1 0.40, Ex2 0.52; distal CSA reduction percentage: Ex1 0.15, Ex2 0.51).
Conclusions: Non-invasive power-mode 3DUS is superior to B-mode 3DUS for imaging and quantification of ICAS. Thereby, further studies are warranted which should now compare power-mode 3DUS with the angiographic gold standard imaging modalities for quantification of ICAS, i.e. with CTA or CE-MRA.
Makouei F, Agander T, Ewertsen C, Svendsen M, Norling R, Kaltoft M J Imaging. 2023; 9(9).
PMID: 37754938 PMC: 10532641. DOI: 10.3390/jimaging9090174.
Recent Advances in Tracking Devices for Biomedical Ultrasound Imaging Applications.
Peng C, Cai Q, Chen M, Jiang X Micromachines (Basel). 2022; 13(11).
PMID: 36363876 PMC: 9695235. DOI: 10.3390/mi13111855.
Spatiotemporal reconstruction method of carotid artery ultrasound from freehand sonography.
Liang H, Ning G, Dai S, Ma L, Luo J, Zhang X Int J Comput Assist Radiol Surg. 2022; 17(9):1731-1743.
PMID: 35704237 DOI: 10.1007/s11548-022-02672-6.
Wang J, Ma Q, Yang Z, Ma L Ann Transl Med. 2020; 8(18):1155.
PMID: 33241004 PMC: 7576085. DOI: 10.21037/atm-20-5713.
de Ruijter J, van Sambeek M, van de Vosse F, Lopata R Med Phys. 2019; 47(3):1034-1047.
PMID: 31837022 PMC: 7079173. DOI: 10.1002/mp.13960.