» Articles » PMID: 11552409

[Toric Intraocular Lenses. Clinical Results and Rotational Stability]

Overview
Journal Ophthalmologe
Specialty Ophthalmology
Date 2001 Sep 13
PMID 11552409
Citations 15
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Correction of an astigmatism at the time of cataract surgery can be achieved in two different ways, by alteration of the corneal curvature or by implantation of a toric intraocular lens (toric IOL). In the latter, in addition to the wound architecture and IOL calculation, the influence of rotational stability in the capsular bag is most important for the refractive result.

Patients And Methods: This retrospective study included 26 eyes from 24 patients with a corneal astigmatism of 2.5-11 D before cataract surgery of which 24 eyes showed a congenital astigmatism and 2 eyes an astigmatism after keratoplasty. After phakoemulsification a three-piece toric PMMA customised IOL (6.5/13.75 mm) was implanted into the capsular bag. Subjective and objective refraction as well as keratometry and corneal topography were performed pre- and postoperatively. The axis of the toric IOL cylinder was marked and could be measured precisely in the postoperative period. The surgically induced astigmatism (SIA) was calculated.

Results: At a mean follow-up time of 12 months after the implantation of a toric IOL, all eyes showed a reduction of total astigmatism. The mean total refractive astigmatism could be reduced from 4.16 D +/- 1.58 D preoperatively to 1.64 D +/- 1.21 D postoperatively. In 6 out of the 26 eyes (23%) the toric IOL rotated more than 10 degrees in the capsular bag and in all 6 cases the IOL rotation happened in the first 3 weeks postoperation. The IOL position was surgically corrected within 3-6 weeks after initial surgery and remained stable during the follow-up period.

Conclusions: The implantation of a PMMA toric IOL is a promising procedure to correct higher levels of corneal astigmatism in cataract surgery. The initial rotational stability of the haptics in the capsular bag still has to be improved and the corneal SIA has to be reduced by smaller incisions. Therefore, three-piece foldable IOLs with a new haptic design are under development.

Citing Articles

Assessment of posture-induced cyclotorsion during cataract surgery using the Verion image-guided system.

Srujana D, Singh R, Titiyal J, Sinha R Med J Armed Forces India. 2021; 77(3):293-296.

PMID: 34305282 PMC: 8282512. DOI: 10.1016/j.mjafi.2020.08.014.


Rotation of a toric intraocular lens with and without capsular tension ring: data from a multicenter non-inferiority randomized clinical trial (RCT).

Hahn U, Krummenauer F, Schmickler S, Koch J BMC Ophthalmol. 2019; 19(1):143.

PMID: 31286913 PMC: 6615103. DOI: 10.1186/s12886-019-1147-5.


[Predictability of residual astigmatism after implantation of posterior chamber toric lenses].

Mies D, Klink T, Eisenbarth W, Meyer L Ophthalmologe. 2017; 115(1):12-20.

PMID: 28289879 DOI: 10.1007/s00347-017-0474-2.


Stability of axis and patient satisfaction after toric implantable collamer lens implantation for myopic astigmatism.

Liu T, Luo X Pak J Med Sci. 2014; 29(6):1371-4.

PMID: 24550956 PMC: 3905361. DOI: 10.12669/pjms.296.3986.


Outcomes and possible risk factors associated with axis alignment and rotational stability after implantation of the Toric implantable collamer lens for high myopic astigmatism.

Sheng X, Rong W, Jia Q, Liu Y, Zhuang W, Gu Q Int J Ophthalmol. 2012; 5(4):459-65.

PMID: 22937505 PMC: 3428541. DOI: 10.3980/j.issn.2222-3959.2012.04.10.