» Articles » PMID: 9922351

Haemolysis Following Implantation of Duct Occlusion Coils

Overview
Journal Heart
Date 1999 Jan 29
PMID 9922351
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: To describe the incidence and management of haemolysis after transcatheter coil occlusion of the arterial duct.

Design: Prospective clinical and echocardiographic follow up of patients who have undergone implantation of the Cook detachable duct occlusion coil.

Setting: Tertiary paediatric cardiac centre.

Patients: Five cases of haemolysis (two girls aged 6 and 11 months; three boys aged 6, 17, and 14 months) from a series of 137 duct coil implantations.

Main Outcome Measures: The occurrence of clinically significant haemolysis after implantation of duct occlusion coils and resolution of haematuria after completion of duct occlusion.

Results: Haemolysis was detected in five of 137 procedures following implantation of Cook detachable duct coils. Four patients became symptomatic 12 hours after the procedure but in one haemolysis was detected three months later. Resolution of ongoing haemolysis was achieved within 48 hours of detection with further coil implantations, but haematuria persisted for up to 10 days. In one patient the extensive destruction of erythrocytes resulted in acute renal failure requiring peritoneal dialysis.

Conclusions: Haemolysis is an important complication after duct coil implantation. It occurred in 3.6% of 137 procedures in this series and is most likely to occur in young patients with relatively large ducts. Further coil implantation to occlude the duct completely is not only successful but technically relatively straightforward and should be undertaken early if major complications such as severe anaemia and renal failure are to be avoided.

Citing Articles

Safety and efficacy of percutaneous patent ductus arteriosus closure: a multicenter Egyptian experience.

Nour A, Abdelrazik Y, Huessin S, Kamel H Egypt Heart J. 2022; 74(1):14.

PMID: 35244792 PMC: 8897530. DOI: 10.1186/s43044-022-00251-3.


Cost-effectiveness analysis of different devices used for the closure of small-to-medium-sized patent ductus arteriosus in pediatric patients.

El-Saiedi S, El Sisi A, Mandour R, Abdel-Aziz D, Attia W Ann Pediatr Cardiol. 2017; 10(2):144-151.

PMID: 28566822 PMC: 5431026. DOI: 10.4103/0974-2069.205138.


Transarterial coil embolization of an abdominal aortocaval fistula in a dog.

Nakata T, Tanaka R, Hamabe L, Yoshiyuki R, Kim S, Suzuki S J Vet Intern Med. 2014; 28(2):656-60.

PMID: 24495213 PMC: 4858001. DOI: 10.1111/jvim.12308.


Complications after transcatheter closure of patent ductus arteriosus.

Jang G, Son C, Lee J, Lee J, Kim S J Korean Med Sci. 2007; 22(3):484-90.

PMID: 17596658 PMC: 2693642. DOI: 10.3346/jkms.2007.22.3.484.


Transcatheter occlusion of the patent ductus arteriosus: a comparison of two devices.

El Mallah M, Sands A, Casey F, Craig B, Mulholland H Ir J Med Sci. 2005; 171(3):151-4.

PMID: 15736355 DOI: 10.1007/BF03170504.

References
1.
Kapoor A, Radhakrishnan S, Shrivastava S . Severe intravascular haemolysis following coil occlusion of patent ductus arteriosus. Indian Heart J. 1996; 48(2):173-4. View

2.
Uzun O, Hancock S, Parsons J, Dickinson D, Gibbs J . Transcatheter occlusion of the arterial duct with Cook detachable coils: early experience. Heart. 1996; 76(3):269-73. PMC: 484519. DOI: 10.1136/hrt.76.3.269. View

3.
Shim D, Wechsler D, Lloyd T, Beekman 3rd R . Hemolysis following coil embolization of a patent ductus arteriosus. Cathet Cardiovasc Diagn. 1996; 39(3):287-90. DOI: 10.1002/(SICI)1097-0304(199611)39:3<287::AID-CCD17>3.0.CO;2-C. View

4.
Henry G, Danilowicz D, Verma R . Severe hemolysis following partial coil-occlusion of patent ductus arteriosus. Cathet Cardiovasc Diagn. 1996; 39(4):410-2. DOI: 10.1002/(SICI)1097-0304(199612)39:4<410::AID-CCD19>3.0.CO;2-G. View

5.
Tometzki A, Arnold R, Peart I, Sreeram N, Abdulhamed J, Godman M . Transcatheter occlusion of the patent ductus arteriosus with Cook detachable coils. Heart. 1996; 76(6):531-5. PMC: 484608. DOI: 10.1136/hrt.76.6.531. View