Ion Recombination Corrections for the NACP Parallel-plate Chamber in a Pulsed Electron Beam

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 1998 Sep 2
PMID 9725587
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

The NACP electron chamber is one of three parallel-plate chambers recommended for use in the UK. Measurements with this chamber type have indicated a problem in determining the recombination correction. This is due to a variation of the ionization current I with polarizing voltage V which deviates from the accepted Boag theory. It is shown that there is a chamber-dependent threshold voltage below which the NACP chamber follows the Boag theory. Above this voltage the chamber should be used with caution, although it is still possible to correct for the dependence of the chamber response on the dose per pulse. The existence of such deviations from theory demonstrates the usefulness of the 1/I against 1/V plot and the limitations of the Boag two-voltage analysis. Values for the initial recombination and the coefficient of general recombination are measured for several NACP chambers. It is shown that from these one can derive a value for the effective plate separation and the collector radius of each chamber. Differences in the behaviour of NACP chambers manufactured by Scanditronix and Dosetek are discussed and the implications of free-electron collection are considered.

Citing Articles

Very High-Energy Electron Therapy Toward Clinical Implementation.

Panaino C, Piccinini S, Andreassi M, Bandini G, Borghini A, Borgia M Cancers (Basel). 2025; 17(2).

PMID: 39857964 PMC: 11763822. DOI: 10.3390/cancers17020181.


Characteristics Performance of Newly Developed Parallel Plate Chamber in Electron Beams.

Saminathan S, Kannan M, Ganesh K, Mazarello J, Fernandes R J Med Phys. 2024; 48(4):333-337.

PMID: 38223789 PMC: 10783180. DOI: 10.4103/jmp.jmp_76_23.


A multicenter study of modified electron beam output calibration.

Mahfirotin D, Ferliano B, Handika A, Asril Y, Fadli M, Ryangga D J Appl Clin Med Phys. 2023; 25(1):e14232.

PMID: 38088260 PMC: 10795448. DOI: 10.1002/acm2.14232.


Determination of the ion collection efficiency of the Razor Nano Chamber for ultra-high dose-rate electron beams.

Cavallone M, Goncalves Jorge P, Moeckli R, Bailat C, Flacco A, Prezado Y Med Phys. 2022; 49(7):4731-4742.

PMID: 35441716 PMC: 9539950. DOI: 10.1002/mp.15675.


Correction for Ion Recombination in a Built-in Monitor Chamber of a Clinical Linear Accelerator at Ultra-High Dose Rates.

Konradsson E, Ceberg C, Lempart M, Blad B, Back S, Knoos T Radiat Res. 2020; 194(6):580-586.

PMID: 33348371 PMC: 7612000. DOI: 10.1667/RADE-19-00012.