In Vitro Accuracy and Reproducibility of Automated and Conventional Periodontal Probes

Overview
Date 1997 May 1
PMID 9178114
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

The aim of this study was to investigate the accuracy and reproducibility of experienced and inexperienced examiners using 3 automated periodontal probes (Florida Pocket Probe, Florida Disk Probe, Peri Probe) in comparison with 3 conventional periodontal probes (Marquis, Williams and EN-15 probes). Test blocks of aluminium had 30 holes of diameter 1.10 mm and depths ranging from 2.75 to 10.0 mm. machined with a tolerance of +/- 0.01 mm. 8 experienced examiners and 8 inexperienced examiners were selected to perform duplicate measurements on the blocks over 6 visits using each of the 6 probes. 1 automated and 1 conventional probe were used at each examination. The % accuracy and reproducibility for each of the duplicate measurements was calculated and analysed using Friedman 2-way analysis of variance and the Wilcoxon matched pairs test. On average, all probes showed high reproducibility, with the Florida Disk Probe, the Florida Pocket Probe and the Williams probe ranked best and the other 3 probes were less reproducible. On average, all probes showed a high degree of accuracy, automated probes were ranked best and were significantly better than conventional probes. Experience had little effect on reproducibility, with only the Peri Probe showing significant differences at the 5% level between the groups. Experience appeared to be more important for accuracy, as experienced examiners were more accurate than inexperienced examiners, with significant differences at the 5% level for the EN-15, Florida Disk Probe and Peri Probe. However, inexperienced examiners were significantly more accurate using the Williams probe. This in vitro study has shown that automated probes offer increased accuracy over conventional probes and the Florida Pocket and disk probes compare well with conventional probes for reproducibility.

Citing Articles

Photoacoustic imaging for monitoring periodontal health: A first human study.

Moore C, Bai Y, Hariri A, Sanchez J, Lin C, Koka S Photoacoustics. 2018; 12:67-74.

PMID: 30450281 PMC: 6226559. DOI: 10.1016/j.pacs.2018.10.005.


Comparative evaluation of probing depth and clinical attachment level using a manual probe and Florida probe.

Kour A, Kumar A, Puri K, Khatri M, Bansal M, Gupta G J Indian Soc Periodontol. 2016; 20(3):299-306.

PMID: 27563204 PMC: 4976551. DOI: 10.4103/0972-124X.181241.


Accuracy and reproducibility of probe forces during simulated periodontal pocket depth measurements.

Al Shayeb K, Turner W, Gillam D Saudi Dent J. 2014; 26(2):50-5.

PMID: 25408596 PMC: 4229682. DOI: 10.1016/j.sdentj.2014.02.001.


In-vitro accuracy and reproducibility evaluation of probing depth measurements of selected periodontal probes.

Al Shayeb K, Turner W, Gillam D Saudi Dent J. 2014; 26(1):19-24.

PMID: 24526824 PMC: 3923167. DOI: 10.1016/j.sdentj.2013.11.001.


The use of povidone-iodine and hydrogen peroxide mixture as an adjunct to non-surgical treatment of slight to moderate chronic periodontitis.

Al-Saeed M, Babay N Saudi Dent J. 2013; 21(3):127-33.

PMID: 23960471 PMC: 3723277. DOI: 10.1016/j.sdentj.2009.10.004.