Evaluation of Respiratory Inductive Plethysmography: Accuracy for Analysis of Respiratory Waveforms

Overview
Journal Chest
Publisher Elsevier
Specialty Pulmonary Medicine
Date 1997 Apr 1
PMID 9106568
Citations 21
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To assess the accuracy of respiratory inductive plethysmography (RIP) waveforms to those obtained with whole body plethysmograph (BP) as this device gives a plethysmographic signal and a pneumotachograph (PNT).

Design: Randomized controlled trial.

Setting: Physiologic laboratory in a university hospital.

Participants: Eleven subjects from the laboratory staff.

Interventions: This study was achieved during four consecutive periods in subjects breathing spontaneously and through different added resistive loads. Using the least square method calibration, two RIP waveforms, VRIP.BP(t) and VRIP.PNT(t), were simultaneously calculated with coefficients obtained from BP and from PNT volume waveforms, respectively VBP(t) and VPNT(t). For each recording, to compare volume waveforms, we calculated their differences in term of distances, DRIP-BP and DRIP-PNT, between the normalized RIP volume signal (respectively, VRIP.BP[t] and VRIP.PNT[t]) and its normalized reference (respectively, VBP[t] and VPNT[t]). We also calculated the distance DPNT-BP between the two normalized references VBP(t) and VPNT(t).

Results: No significant effect of load or time on the distance occurred. Including all the recordings, the mean distance DRIP-BP (3.4+/-1.1%) appears significantly lower than both the mean distance DRIP-PNT (4.5+/-1.3%; p<0.04) and the mean distance DPNT-BP (4.6+/-0.9%; p<0.008). For each period or load level, DRIP-BP appears to be lower than DRIP-PNT and DPNT-BP.

Conclusion: The RIP seems reasonably accurate for analysis of respiratory waveform while subjects subsequently breathe against resistive loads.

Citing Articles

Slower respiration rate is associated with higher self-reported well-being after wellness training.

Kral T, Weng H, Mitra V, Imhoff-Smith T, Azemi E, Goldman R Sci Rep. 2023; 13(1):15953.

PMID: 37743388 PMC: 10518325. DOI: 10.1038/s41598-023-43176-w.


Wearable Vital Signs Monitoring for Patients With Asthma: A Review.

Taylor L, Ding X, Clifton D, Lu H IEEE Sens J. 2023; 23(3):1734-1751.

PMID: 37655115 PMC: 7615004. DOI: 10.1109/JSEN.2022.3224411.


Wearable Fabric Loop Sensor Based on Magnetic-Field-Induced Conductivity for Simultaneous Detection of Cardiac Activity and Respiration Signals.

Cho H, Yang J, Lee S, Lee J, Lee J Sensors (Basel). 2022; 22(24).

PMID: 36560254 PMC: 9786050. DOI: 10.3390/s22249884.


The influence of receiving real-time visual feedback on breathing during treadmill running to exhaustion.

Passafiume J, Glover N, Crecelius A, Chaudhari A PLoS One. 2022; 17(7):e0270335.

PMID: 35797373 PMC: 9262203. DOI: 10.1371/journal.pone.0270335.


Measurement of chest wall motion using a motion capture system with the one-pitch phase analysis method.

Tamiya H, Mitani A, Isago H, Ishimori T, Saito M, Jo T Sci Rep. 2021; 11(1):21497.

PMID: 34728740 PMC: 8563798. DOI: 10.1038/s41598-021-01033-8.