Abstract

In a canine puppy model, pulmonary artery stenosis was created by banding the left pulmonary artery to 30-40% of its original diameter. Animals underwent right heart catheterization and angiography 1-2 mo later, and Palmaz P308 stents were implanted. Stent redilation was performed 3-5 mo later. One mo postredilation, the animals were restudied and sacrificed. Coarctations of the aorta were created by transverse aortic incision and longitudinal repair. P308 stent implantation was performed 2-3 mo later. Stent redilation was performed after 6-10 mo, and the animals were restudied and sacrificed 1-2 mo later. Stent implantation was performed in 6 puppies with pulmonary artery stenosis, as 2 animals developed postoperative pulmonary arterial hypoplasia, precluding stenting. The stenosis diameter increased from 4.8 +/- 0.5 mm to 7.4 +/- 0.6 mm (mean +/- SE) following stenting (P = 0.005), and increased further to 9.2 +/- 0.7 mm following redilation (P < 0.001). There were no significant vessel tears or ruptures. Coarctation stenting was performed in 8 animals. The coarctation was dilated from 5.8 +/- 0.9 mm to 9.8 +/- 0.6 mm (P < 0.001), and to 13.5 +/- 0.5 mm at redilation (P = 0.002). Redilation could not be performed in 1 animal. Aortic rupture and death occurred in 2 of 7 animals at redilation. Stent implantation and redilation in experimental pulmonary artery stenosis appears safe and effective. Though stent implantation for coarctation of the aorta appears safe, there was a 28% aortic rupture rate at stent redilation in this model.

Citing Articles

Balloon expandable stents for coarctation of the aorta: review of current status and technical considerations.

Ebeid M Images Paediatr Cardiol. 2012; 5(2):25-41.

PMID: 22368626 PMC: 3232539.


Cost-effectiveness of coarctation repair strategies: endovascular stenting versus surgery.

George J, Shim D, Bucuvalas J, Immerman E, Manning P, Pearl J Pediatr Cardiol. 2003; 24(6):544-7.

PMID: 12881774 DOI: 10.1007/s00246-003-0496-1.


The efficacy and safety of stent redilatation in congenital heart disease.

Duke C, Rosenthal E, Qureshi S Heart. 2003; 89(8):905-12.

PMID: 12860870 PMC: 1767765. DOI: 10.1136/heart.89.8.905.


Stent implantation for aortic coarctation and recoarctation.

Magee A, Qureshi S, Rosenthal E, Zubrzycka M, Ksiazyk J, Tynan M Heart. 1999; 82(5):600-6.

PMID: 10525517 PMC: 1760772. DOI: 10.1136/hrt.82.5.600.