Accuracy and Precision Study in Vitro for Peripheral Quantitative Computed Tomography

Overview
Journal Osteoporos Int
Date 1996 Jan 1
PMID 8783294
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

We evaluated the accuracy and precision of a peripheral quantitative computed tomography (pQCT) scanner, the Stratec XCT-960, using 12 human cadaveric forearms. The accuracy was determined by comparing the total bone mineral content (BMC) with the ash weight (AW). We scanned and ashed three consecutive slices (thickness 2.5 mm) at the standard position (s-position) and at 2.5 mm both proximal and distal to the s-position. The correlation coefficient between the AW and total BMC using slices at the s-position was r = 0.87 with an accuracy error (random component) of 15.5%. The correlation coefficient using all slices was r = 0.90 with an accuracy error of 14.3%. The correlation coefficient improved to r = 0.95 with an accuracy error of 9.7% after averaging the results of all three slices for each forearm. The short-term precision error expressed as the coefficient of variation (CV) of bone mineral density (BMD) and BMC was determined by measuring the forearms five times either with repositioning or without repositioning. The CVs with repositioning were 2.77 and 1.15 for total BMD and BMC, 1.85 for trabecular BMD; without repositioning they were 0.29, 0.58 and 0.69 respectively. To further evaluate the influence of positioning, additional scans were performed at 1, 2 and 5 mm proximal, and 1 and 2 mm distal to the s-position. BMD and BMC were greatly influenced by the scan location; for example, the percentage differences in trabecular BMD 1 mm distal and proximal relative to the s-position were 2.5% +/- 5.1% and 0.18% +/- 6.3%, respectively. The Stratec XCT-960 appears to be a moderately accurate and highly precise scanner with potential usefulness for evaluating BMC and BMD of ultradistal radius.

Citing Articles

Estimation of whole body fat from appendicular soft tissue from peripheral quantitative computed tomography in adolescent girls.

Lee V, Blew R, Farr J, Tomas R, Lohman T, Going S Int J Body Compos Res. 2014; 11(1):1-8.

PMID: 25147482 PMC: 4137877.


Reproducibility of peripheral quantitative computed tomography measurements at the radius and tibia in healthy pre- and postmenopausal women.

Szabo K, Webber C, Gordon C, Adachi J, Tozer R, Papaioannou A Can Assoc Radiol J. 2010; 62(3):183-9.

PMID: 20576400 PMC: 5101073. DOI: 10.1016/j.carj.2010.04.011.


The impact of accurate positioning on measurements made by peripheral QCT in the distal radius.

Marjanovic E, Ward K, Adams J Osteoporos Int. 2008; 20(7):1207-14.

PMID: 18982400 DOI: 10.1007/s00198-008-0778-9.


Well-nourished cystic fibrosis patients have normal mineral density, but reduced cortical thickness at the forearm.

Louis O, Clerinx P, Gies I, De Wachter E, De Schepper J Osteoporos Int. 2008; 20(2):309-14.

PMID: 18536951 DOI: 10.1007/s00198-008-0646-7.


Methods for measurement of pediatric bone.

Binkley T, Berry R, Specker B Rev Endocr Metab Disord. 2008; 9(2):95-106.

PMID: 18247119 DOI: 10.1007/s11154-008-9073-5.


References
1.
Gluer C, Reiser U, Davis C, Rutt B, Genant H . Vertebral mineral determination by quantitative computed tomography (QCT): accuracy of single and dual energy measurements. J Comput Assist Tomogr. 1988; 12(2):242-58. DOI: 10.1097/00004728-198803000-00013. View

2.
LeBoff M, Fuleihan G, Angell J, Chung S, Curtis K . Dual-energy x-ray absorptiometry of the forearm: reproducibility and correlation with single-photon absorptiometry. J Bone Miner Res. 1992; 7(7):841-6. DOI: 10.1002/jbmr.5650070714. View

3.
Nelson D, Feingold M, Mascha E, Kleerekoper M . Comparison of single-photon and dual-energy x-ray absorptiometry of the radius. Bone Miner. 1992; 18(1):77-83. DOI: 10.1016/0169-6009(92)90801-j. View

4.
Hagiwara S, Engelke K, Yang S, Dhillon M, Guglielmi G, Nelson D . Dual x-ray absorptiometry forearm software: accuracy and intermachine relationship. J Bone Miner Res. 1994; 9(9):1425-7. DOI: 10.1002/jbmr.5650090914. View

5.
Faulkner K, McClung M, Schmeer M, Roberts L, Gaither K . Densitometry of the radius using single and dual energy absorptiometry. Calcif Tissue Int. 1994; 54(3):208-11. DOI: 10.1007/BF00301680. View