Radioactive Microsphere Validation of a Volume Localized Continuous Saturation Perfusion Measurement

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 1994 Feb 1
PMID 8133750
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

Detre et al. (Magn. Reson. Med. 23, 37-45 (1992)) and Zhang et al., (Magn. Reson. Med. 25, 362 (1992)) have recently demonstrated a technique for the measurement of regional cerebral blood flow (rCBF) based on the continuous saturation (or inversion) of the arterial blood supply to the brain in rats at 4.7 T. In the work reported here, we combined this technique with volume localized (PRESS) readouts to benefit from recording "perfusion" signals averaged over a larger volume, resulting in rapid acquisition of data with sufficient signal-to-noise ratio for application at 2.0 T. In 10 baseline flow measurements, the mean error between the NMR technique and the microsphere flow measurement was -1.5% with a standard deviation of 15.2%. For five measurements obtained with occlusion of the middle cerebral artery, the mean error was -32.4 +/- 20.2%. Perfusion measurements from a single animal under hypercapnic conditions indicated that the NMR technique could underestimate rCBF at high flow rates. An error analysis of the NMR perfusion model is also presented, along with results for typical parameters encountered at 2.0 T.

Citing Articles

Tumor Vascularity in Renal Masses: Correlation of Arterial Spin-Labeled and Dynamic Contrast-Enhanced Magnetic Resonance Imaging Assessments.

Zhang Y, Kapur P, Yuan Q, Xi Y, Carvo I, Signoretti S Clin Genitourin Cancer. 2015; 14(1):e25-36.

PMID: 26422014 PMC: 4698181. DOI: 10.1016/j.clgc.2015.08.007.


Pain response measured with arterial spin labeling.

Maleki N, Brawn J, Barmettler G, Borsook D, Becerra L NMR Biomed. 2013; 26(6):664-73.

PMID: 23319440 PMC: 3634868. DOI: 10.1002/nbm.2911.


Spatial-temporal dynamics of pulmonary blood flow in the healthy human lung in response to altered FI(O2).

Asadi A, Cronin M, Sa R, Theilmann R, Holverda S, Hopkins S J Appl Physiol (1985). 2012; 114(1):107-18.

PMID: 23104691 PMC: 3544514. DOI: 10.1152/japplphysiol.00433.2012.


Cerebral blood flow quantification in the rat: a direct comparison of arterial spin labeling MRI with radioactive microsphere PET.

Bos A, Bergmann R, Strobel K, Hofheinz F, Steinbach J, van den Hoff J EJNMMI Res. 2012; 2(1):47.

PMID: 22978819 PMC: 3598883. DOI: 10.1186/2191-219X-2-47.


Arterial Spin Labeling (ASL) fMRI: advantages, theoretical constrains, and experimental challenges in neurosciences.

Borogovac A, Asllani I Int J Biomed Imaging. 2012; 2012:818456.

PMID: 22966219 PMC: 3432878. DOI: 10.1155/2012/818456.