Accuracy of Fingerstick Glucose Values in Shock Patients

Overview
Journal Am J Crit Care
Specialties Critical Care
Nursing
Date 1995 Jan 1
PMID 7894555
Citations 22
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Fingerstick blood glucose measurement has become widespread in both hospital and prehospital settings.

Objective: To determine the accuracy of fingerstick blood glucose measurement in patients with poor peripheral perfusion (shock).

Method: Results obtained during three methods of glucose analysis (fingerstick blood glucose measurement; bedside and laboratory glucose analysis) were examined prospectively on 38 patients from inpatient medical and surgical critical care units or the emergency department of a large tertiary care referral center.

Results: The means of the three glucose measurements were significantly different. Univariate analysis of the mean laboratory glucose value versus the mean fingerstick glucose value was significantly different. The mean venipuncture glucose measured by the bedside glucose meter versus the mean venous laboratory glucose was not significantly different.

Conclusion: These results suggest that fingerstick blood samples should not be used for bedside glucose analysis in patients who may have inadequate tissue perfusion.

Citing Articles

Semi-Implantable Bioelectronics.

Fang J, Huang S, Liu F, He G, Li X, Huang X Nanomicro Lett. 2022; 14(1):125.

PMID: 35633391 PMC: 9148344. DOI: 10.1007/s40820-022-00818-4.


Comparing Self Monitoring Blood Glucose Devices and Laboratory Tests: Over 25 Years Experience.

Harada Y, Harada K, Chin Jr P Cureus. 2020; 11(12):e6268.

PMID: 31903305 PMC: 6937471. DOI: 10.7759/cureus.6268.


Reduction of measurement noise in a continuous glucose monitor by coating the sensor with a zwitterionic polymer.

Xie X, Doloff J, Yesilyurt V, Sadraei A, McGarrigle J, Omami M Nat Biomed Eng. 2019; 2(12):894-906.

PMID: 30931173 PMC: 6436621. DOI: 10.1038/s41551-018-0273-3.


Accuracy of Capillary and Arterial Whole Blood Glucose Measurements Using a Glucose Meter in Patients under General Anesthesia in the Operating Room.

Karon B, Donato L, Larsen C, Siebenaler L, Wells A, Wood-Wentz C Anesthesiology. 2017; 127(3):466-474.

PMID: 28557817 PMC: 5561475. DOI: 10.1097/ALN.0000000000001708.


Glucose Meter Use in the Intensive Care Unit: Much Ado About Something.

Brad S EJIFCC. 2016; 25(2):178-87.

PMID: 27683465 PMC: 4975293.