Accuracy of Capillary and Arterial Whole Blood Glucose Measurements Using a Glucose Meter in Patients Under General Anesthesia in the Operating Room

Overview
Journal Anesthesiology
Specialty Anesthesiology
Date 2017 May 31
PMID 28557817
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The aim of this study was to evaluate the use of a glucose meter with surgical patients under general anesthesia in the operating room.

Methods: Glucose measurements were performed intraoperatively on 368 paired capillary and arterial whole blood samples using a Nova StatStrip (Nova Biomedical, USA) glucose meter and compared with 368 reference arterial whole blood glucose measurements by blood gas analyzer in 196 patients. Primary outcomes were median bias (meter minus reference), percentage of glucose meter samples meeting accuracy criteria for subcutaneous insulin dosing as defined by Parkes error grid analysis for type 1 diabetes mellitus, and accuracy criteria for intravenous insulin infusion as defined by Clinical and Laboratory Standards Institute guidelines. Time under anesthesia, patient position, diabetes status, and other variables were studied to determine whether any affected glucose meter bias.

Results: Median bias (interquartile range) was -4 mg/dl (-9 to 0 mg/dl), which did not differ from median arterial meter bias of -5 mg/dl (-9 to -1 mg/dl; P = 0.32). All of the capillary and arterial glucose meter values met acceptability criteria for subcutaneous insulin dosing, whereas only 89% (327 of 368) of capillary and 93% (344 of 368) arterial glucose meter values met accuracy criteria for intravenous insulin infusion. Time, patient position, and diabetes status were not associated with meter bias.

Conclusions: Capillary and arterial blood glucose measured using the glucose meter are acceptable for intraoperative subcutaneous insulin dosing. Whole blood glucose on the meter did not meet accuracy guidelines established specifically for more intensive (e.g., intravenous insulin) glycemic control in the acute care environment.

Citing Articles

Diagnostic agreement between three point-of-care glucose and β-hydroxybutyrate meters and reference laboratory methods in stingrays.

Dannemiller N, Cray C, Westmoreland L, Christiansen E Front Vet Sci. 2024; 10:1254340.

PMID: 38173551 PMC: 10761536. DOI: 10.3389/fvets.2023.1254340.


Accuracy of the Nova StatStrip glucometer in patients undergoing major abdominal surgery: an observational study.

Omiya K, Nakadate Y, Sato H, Koo B, Schricker T Can J Anaesth. 2023; 70(12):1970-1977.

PMID: 37919630 DOI: 10.1007/s12630-023-02606-z.


What, why and how to monitor blood glucose in critically ill patients.

Juneja D, Deepak D, Nasa P World J Diabetes. 2023; 14(5):528-538.

PMID: 37273246 PMC: 10236998. DOI: 10.4239/wjd.v14.i5.528.


Perioperative challenges in management of diabetic patients undergoing non-cardiac surgery.

Galway U, Chahar P, Schmidt M, Araujo-Duran J, Shivakumar J, Turan A World J Diabetes. 2021; 12(8):1255-1266.

PMID: 34512891 PMC: 8394235. DOI: 10.4239/wjd.v12.i8.1255.


[Treatment of diabetes mellitus in perioperative medicine-an update].

Rupprecht B, Stockl A, Stockl S, Dietrich C Anaesthesist. 2020; 70(6):451-465.

PMID: 33141238 DOI: 10.1007/s00101-020-00875-7.


References
1.
Dhatariya K, Levy N, Kilvert A, Watson B, Cousins D, Flanagan D . NHS Diabetes guideline for the perioperative management of the adult patient with diabetes. Diabet Med. 2012; 29(4):420-33. DOI: 10.1111/j.1464-5491.2012.03582.x. View

2.
Herasevich V, Kor D, Li M, Pickering B . ICU data mart: a non-iT approach. A team of clinicians, researchers and informatics personnel at the Mayo Clinic have taken a homegrown approach to building an ICU data mart. Healthc Inform. 2011; 28(11):42, 44-5. View

3.
Wiener R, Wiener D, Larson R . Benefits and risks of tight glucose control in critically ill adults: a meta-analysis. JAMA. 2008; 300(8):933-44. DOI: 10.1001/jama.300.8.933. View

4.
Charlson M, Szatrowski T, Peterson J, Gold J . Validation of a combined comorbidity index. J Clin Epidemiol. 1994; 47(11):1245-51. DOI: 10.1016/0895-4356(94)90129-5. View

5.
Boyd J, Bruns D . Effects of measurement frequency on analytical quality required for glucose measurements in intensive care units: assessments by simulation models. Clin Chem. 2014; 60(4):644-50. DOI: 10.1373/clinchem.2013.216366. View