A new apparatus has been developed using the laser speckle phenomenon for non-contact, two-dimensional analysis of tissue circulation in the choroid and optic nerve head (ONH). The fundus was illuminated by a diode laser spot and its image speckle was detected by an image sensor. The difference between the average of the speckle intensity (Imean) and the speckle intensity for successive scannings was calculated, and the ratio of Imean to this difference was defined as normalized blur (NB), which is a quantitative index of blood velocity. The results were displayed on color graphic monitors showing the two-dimensional variation of the NB level in the measurement field. In the rabbit, this apparatus was used to study the relationship between the results of NB measurement and the choroidal blood flow determined by the microsphere technique, the relationship between NB obtained from the ONH tissue free of visible surface vessels and the ocular perfusion pressure (OPP) after a lethal injection of pentobarital, and the effect of intraocular pressure (IOP) on the NB in the choroid or in the ONH. A stepwise reduction in the OPP was introduced by elevating the IOP manometrically. The relative decrease in the average NB over the field measured, NBav, in the choroid with the reduction in OPP showed a significant correlation with the relative change in the blood flow rate determined using the microsphere technique (r = 0.60, P < 0.001). NBav in the ONH had a good correlation with the OPP after a lethal injection of pentobarbital (r = 0.98, P < 0.001). NBav in the choroid decreased with reduction in the OPP. Although NBav in the ONH was little affected by OPP change when OPP was above 40 mmHg, at OPP levels below 40 mmHg, NBav in the ONH decreased along with a reduction in the OPP. These results suggest that by using the present apparatus, the blood velocity in the choroid or ONH under various conditions can be studied non-invasively in the living eye.

Citing Articles

Multimodal Eye Imaging, Retina Characteristics, and Psychological Assessment Dataset.

Zhang G, Qu Y, Zhang Y, Tang J, Wang C, Yin H Sci Data. 2024; 11(1):836.

PMID: 39095400 PMC: 11297319. DOI: 10.1038/s41597-024-03690-6.


Evaluating the relationship between ocular blood flow and systemic organ blood flow in hemorrhagic shock using a rabbit model.

Watanabe K, Shiba T, Takahara A, Homma H, Komatsu T, Tanino Y Sci Rep. 2024; 14(1):3749.

PMID: 38355984 PMC: 10866860. DOI: 10.1038/s41598-024-54467-1.


Improvement of clinical wound microcirculation diagnosis using an object tracking-based laser speckle contrast imaging system.

Hsieh M, Chang C, Hsu C, Lin Y, Hsieh P, Ching C APL Bioeng. 2024; 8(1):016105.

PMID: 38292062 PMC: 10827336. DOI: 10.1063/5.0172443.


Novel superpixel method to visualize fundus blood flow resistivity in healthy adults.

Okamoto K, Takahashi N, Kobayashi T, Shiba T, Hori Y, Fujii H Sci Rep. 2023; 13(1):6171.

PMID: 37061579 PMC: 10105763. DOI: 10.1038/s41598-023-33450-2.


Ocular blood flow evaluation by laser speckle flowgraphy in pediatric patients with anisometropia.

Itokawa T, Matsumoto T, Matsumura S, Kawakami M, Hori Y Front Public Health. 2023; 11:1093686.

PMID: 36923046 PMC: 10010384. DOI: 10.3389/fpubh.2023.1093686.