A Test Pattern for Quality Control of Laser Scanner and Charge-coupled Device Film Digitizers

Overview
Journal J Digit Imaging
Publisher Springer
Date 1995 Feb 1
PMID 7734527
Citations 2
Authors
Affiliations
Soon will be listed here.
Abstract

Although clinical images provide the ultimate test of diagnostic performance for a film digitizer, such images are not generally suitable for daily quality control (QC) purposes. However, a well-designed test pattern will provide a rapid, comprehensive, objective and reproducible assessment of image quality. This pattern should evaluate various parameters of image quality, including high contrast resolution, low contrast discrimination, linearity of gray scale, geometric distortion, and noise. Furthermore, the pattern should detect light leak and film slippage, two problems commonly associated with film digitizers. The test pattern described in this manuscript was designed to provide quantitative measures of performance for a film digitizer. As part of a regular QC routine for a laser scanner or charge-coupled device digitizer, this pattern provides a simple method to identify and quantify changes in digital image quality.

Citing Articles

The laser film digitizer: density, contrast, and resolution.

Teslow T J Digit Imaging. 1997; 10(3 Suppl 1):128-32.

PMID: 9268857 PMC: 3452839. DOI: 10.1007/BF03168675.


Quality assurance and quality control of an intensive care unit picture archiving and communication system.

Tucker D, McEachern M J Digit Imaging. 1995; 8(4):162-7.

PMID: 8573625 DOI: 10.1007/BF03168715.

References
1.
Good W, Gur D, FEIST J, Thaete F, Fuhrman C, Britton C . Subjective and objective assessment of image quality--a comparison. J Digit Imaging. 1994; 7(2):77-8. DOI: 10.1007/BF03168426. View

2.
Ackerman S, GITLIN J, Gayler R, FLAGLE C, Bryan R . Receiver operating characteristic analysis of fracture and pneumonia detection: comparison of laser-digitized workstation images and conventional analog radiographs. Radiology. 1993; 186(1):263-8. DOI: 10.1148/radiology.186.1.8416576. View

3.
Roehrig H, Blume H, Ji T, Browne M . Performance tests and quality control of cathode ray tube displays. J Digit Imaging. 1990; 3(3):134-45. DOI: 10.1007/BF03167599. View

4.
Halpern E, Esser P . An improved phantom for quality control of laser scanner digitizers in picture archival and communications systems. J Digit Imaging. 1991; 4(4):241-7. DOI: 10.1007/BF03173906. View

5.
Slasky B, Gur D, Good W, Harris K, Cooperstein L, Rockette H . Receiver operating characteristic analysis of chest image interpretation with conventional, laser-printed, and high-resolution workstation images. Radiology. 1990; 174(3 Pt 1):775-80. DOI: 10.1148/radiology.174.3.2305061. View