P0.1 is a Useful Parameter in Setting the Level of Pressure Support Ventilation

Overview
Specialty Critical Care
Date 1995 Jul 1
PMID 7593895
Citations 36
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: The purpose of this study was to investigate whether changes in breathing pattern, neuromuscular drive (P0.1), and the work involved in breathing might help to set the individual appropriate level of pressure support ventilation (PSV) in patients with acute respiratory failure (ARF) requiring ventilatory assistance.

Design: A prospective, interventional study.

Setting: An 8-bed multidisciplinary intensive care unit (ICU).

Patients: Ten patients with ARF due to adult respiratory distress syndrome (ARDS), sepsis or airway infection were included in the study. Chronic obstructive pulmonary disease (COPD) patients with acute exacerbation were excluded. None of these patients was in the weaning process.

Interventions: We found a level of pressure support able to generate a condition of near-relaxation in each patient, as evidenced by work of breathing (WOB) values close to 0 J/1. This level was called PS 100 and baseline physiological measurements, namely, breathing pattern, P0.1 and WOB were obtained. Pressure support was then reduced to 85%, 70%, and 50% of the initial value and the same set of measurements was obtained.

Measurements And Results: Flow (V) was measured by a flow sensor (Varflex) positioned between the Y-piece of the breathing circuit and the endotracheal tube. Tidal volume was obtained by numerical integration of the flow signal. Airway pressure (P(aw)) was sampled through a catheter attached to the flow sensor. Esophageal pressure (Pes) was measured with a nasogastric tube incorporating an esophageal balloon. The esophageal balloon and flow and pressure sensors were connected to a portable monitor (CP 100 Bicore) that provided real-time display of flow, volume, P(aw) and Pes tracings and loops of Pes/V, P(aw)/V and V/V relationships. The breathing pattern was analyzed from the flow signal. Patient work of breathing (WOB) was calculated by integration of the area of the Pes/V loop. Respiratory drive (P0.1) was measured at the esophageal pressure change during the first 100 ms of a breath, by the quasi-occlusion technique. When pressure support was reduced, we found that the respiration rate significantly increased from PS 100 to PS 85, but varied negligibly with lower pressure support levels. Tidal volume behaved in a similar way, decreasing significantly from PS 100 to PS 85, but hardly changing at PS 70 and PS 50. In contrast, WOB and P0.1 increased progressively with decreasing pressure support levels. The changes in WOB were significant at each stage in the trial, whereas P0.1 increased significantly from PS 100 at other stages. Linear regression analysis revealed a highly positive, significant correlation between WOB and P0.1 at decreasing PSV levels (r = 0.87), whereas the correlation between WOB and ventilatory frequency was less significant (r = 0.53). No other correlation was found.

Conclusions: During pressure support ventilation, P0.1 may be a more sensitive parameter than the assessment of breathing pattern in setting the optimal level of pressure support in individual patients. Although P0.1 was measured with an esophageal balloon in the present study, non-invasive techniques can also be used.

Citing Articles

Ventilator performances for non-invasive ventilation: a bench study.

Caillard C, Fresnel E, Artaud-Macari E, Cuvelier A, Tamion F, Patout M BMJ Open Respir Res. 2024; 11(1).

PMID: 39438080 PMC: 11499821. DOI: 10.1136/bmjresp-2023-002144.


Respiratory Drive, Effort, and Lung-Distending Pressure during Transitioning from Controlled to Spontaneous Assisted Ventilation in Patients with ARDS: A Multicenter Prospective Cohort Study.

Balzani E, Murgolo F, Pozzi M, Di Mussi R, Bartolomeo N, Simonetti U J Clin Med. 2024; 13(17).

PMID: 39274439 PMC: 11396025. DOI: 10.3390/jcm13175227.


Positive end-expiratory pressure limits inspiratory effort through modulation of the effort-to-drive ratio: an experimental crossover study.

Widing H, Pellegrini M, Chiodaroli E, Persson P, Hallen K, Perchiazzi G Intensive Care Med Exp. 2024; 12(1):10.

PMID: 38311676 PMC: 10838888. DOI: 10.1186/s40635-024-00597-9.


Use of airway pressure-based indices to detect high and low inspiratory effort during pressure support ventilation: a diagnostic accuracy study.

Yang Y, Liu Y, Gao R, Song D, Zhou Y, Miao M Ann Intensive Care. 2023; 13(1):111.

PMID: 37955842 PMC: 10643759. DOI: 10.1186/s13613-023-01209-7.


ERS International Congress 2022: highlights from the Respiratory Intensive Care Assembly.

Valentin S, Lopez Padilla D, Nolasco S, Ranilovic D, Guillamat-Prats R, Marin T ERJ Open Res. 2023; 9(3).

PMID: 37228293 PMC: 10204864. DOI: 10.1183/23120541.00532-2022.


References
1.
Marini J . Strategies to minimize breathing effort during mechanical ventilation. Crit Care Clin. 1990; 6(3):635-61. View

2.
Macintyre N . Respiratory function during pressure support ventilation. Chest. 1986; 89(5):677-83. DOI: 10.1378/chest.89.5.677. View

3.
Nathan S, Ishaaya A, KOERNER S, Belman M . Prediction of minimal pressure support during weaning from mechanical ventilation. Chest. 1993; 103(4):1215-9. DOI: 10.1378/chest.103.4.1215. View

4.
Brochard L, Pluskwa F, Lemaire F . Improved efficacy of spontaneous breathing with inspiratory pressure support. Am Rev Respir Dis. 1987; 136(2):411-5. DOI: 10.1164/ajrccm/136.2.411. View

5.
Henning R, SHUBIN H, Weil M . The measurement of the work of breathing for the clinical assessment of ventilator dependence. Crit Care Med. 1977; 5(6):264-8. DOI: 10.1097/00003246-197711000-00002. View