Breath-by-breath P0.1 Measured on Quasi-occlusion Via Hamilton C6 May Result in Underestimation of Respiratory Drive and Inspiratory Effort
Overview
Authors
Affiliations
We aimed to identify the threshold for P0.1 in a breath-by-breath manner measured by the Hamilton C6 on quasi-occlusion for high respiratory drive and inspiratory effort. In this prospective observational study, we analyzed the relationships between airway P0.1 on quasi-occlusion and esophageal pressure (esophageal P0.1 and esophageal pressure swing). We also conducted a linear regression analysis and derived the threshold of airway P0.1 on quasi-occlusion for high respiratory drive and inspiratory effort. We found that airway P0.1 measured on quasi-occlusion had a strong positive correlation with esophageal P0.1 measured on quasi-occlusion and esophageal pressure swing, respectively. Additionally, the P0.1 threshold for high respiratory drive and inspiratory effort were calculated at approximately 1.0 cmHO from the regression equations. Our calculations suggest a lower threshold of airway P0.1 measured by the Hamilton C6 on quasi-occlusion than that which has been previously reported.
The Paradox of Occlusion Pressure at 0.1 s (P) Measurement Without Airway Occlusion.
Iotti G, Negri M, Schranz C Respir Care. 2024; 69(4):524-525.
PMID: 38538013 PMC: 11108113. DOI: 10.4187/respcare.11813.
Ito Y, Herrera M, Hotz J, Kyogoku M, Newth C, Bhalla A Crit Care. 2023; 27(1):466.
PMID: 38031116 PMC: 10685539. DOI: 10.1186/s13054-023-04754-6.