Kinematics of Successful Knee Prostheses During Weight-bearing: Three-dimensional Movements and Positions of Screw Axes in the Tricon-M and Miller-Galante Designs

Overview
Publisher Wiley
Date 1994 Jan 1
PMID 7584178
Citations 17
Authors
Affiliations
Soon will be listed here.
Abstract

Using roentgen stereophotogrammetric analysis we recorded the three-dimensional movements in six knees with implanted Tricon-M prostheses and ten knees with Miller-Galante prostheses as the patients ascended a platform. Fourteen patients with normal knees were used as controls. The two prosthetic designs displayed decreased internal tibial rotation and the Tricon-M increased valgus rotation. A central point on the tibial articular surface had a more lateral position in the Tricon-M design and a more distal one in the Miller-Galante design compared to normal knees. Increased posterior displacement with increasing flexion was observed in both designs. When the normal knees were extended at full weight-bearing the helical axes mainly shifted inclination in the frontal plane. In the prosthetic knees there was a tendency to anterior-posterior displacement of the axes as extension proceeded, especially in the Miller-Galante design. Translations along the helical axes were larger than normal in the Miller-Galante and smaller in the Tricon-M knees, reflecting differences in constraint of the two designs.

Citing Articles

Dynamic knee control and movement strategies in athletes and non-athletes in side hops: Implications for knee injury.

Markstrom J, Grip H, Schelin L, Hager C Scand J Med Sci Sports. 2019; 29(8):1181-1189.

PMID: 30972848 PMC: 6850355. DOI: 10.1111/sms.13432.


Osteophytes removal in patella versus lateral facetectomy of the patella in patella-retaining total knee arthroplasty.

Moghtadaei M, Sajadi M, Farahini H, Yeganeh A Interv Med Appl Sci. 2017; 9(1):15-19.

PMID: 28932491 PMC: 5598117. DOI: 10.1556/1646.9.2017.1.07.


Kinematic analysis of posterior-stabilized total knee arthroplasty during standing up from and sitting down on a chair.

Mine T, Hoshi K, Gamada K, Ihara K, Kawamura H, Kuriyama R J Orthop Surg Res. 2016; 11(1):142.

PMID: 27855716 PMC: 5114786. DOI: 10.1186/s13018-016-0482-y.


In vivo kinematics of a unique posterior-stabilized knee implant during a stepping exercise.

Mine T, Hoshi K, Gamada K, Ihara K, Kawamura H, Kuriyama R J Orthop Surg Res. 2016; 11:18.

PMID: 26831568 PMC: 4736098. DOI: 10.1186/s13018-016-0354-5.


Influence of mediolateral tibial baseplate position in TKA on knee kinematics and retropatellar pressure.

Steinbruck A, Fottner A, Schroder C, Woiczinski M, Schmitt-Sody M, Muller T Knee Surg Sports Traumatol Arthrosc. 2015; 25(8):2602-2608.

PMID: 26531185 DOI: 10.1007/s00167-015-3843-x.


References
1.
Walker P, Shoji H, Erkman M . The rotational axis of the knee and its significance to prosthesis design. Clin Orthop Relat Res. 1972; 89:160-70. View

2.
Walker P, HAJEK J . The load-bearing area in the knee joint. J Biomech. 1972; 5(6):581-9. DOI: 10.1016/0021-9290(72)90030-9. View

3.
Dorr L, Ochsner J, Gronley J, Perry J . Functional comparison of posterior cruciate-retained versus cruciate-sacrificed total knee arthroplasty. Clin Orthop Relat Res. 1988; (236):36-43. View

4.
Frankel V, Burstein A, Brooks D . Biomechanics of internal derangement of the knee. Pathomechanics as determined by analysis of the instant centers of motion. J Bone Joint Surg Am. 1971; 53(5):945-62. View

5.
Soudan K, Van Audekercke R, Martens M . Exact knowledge of knee motion kinematics as a basis for well-designed knee prostheses : the instant axis concept. Acta Orthop Belg. 1980; 46(6):757-65. View