The Use of Quartz Patch Pipettes for Low Noise Single Channel Recording

Overview
Journal Biophys J
Publisher Cell Press
Specialty Biophysics
Date 1993 Oct 1
PMID 7506069
Citations 34
Authors
Affiliations
Soon will be listed here.
Abstract

Quartz has a dissipation factor of approximately 10(-4), which is an order of magnitude less than that of the best glasses previously used to fabricate patch pipettes; it's dielectric constant of 3.8 is also lower than that of other glasses. On the basis of these electrical characteristics it is expected that patch pipettes pulled from quartz tubing will produce significantly less noise than pipettes made from other glasses. Our work confirms these expectations and we describe theoretical and practical aspects of the use of quartz pipettes for single channel patch voltage clamp measurements. Methods for pulling quartz pipettes with a laser-based puller and coating them with low-loss elastomers are discussed, as are precautions that are necessary to achieve low noise recordings. We have shown that quartz pipettes can be pulled from tubing with outer diameter to inner diameter ratios as large as 3 and a method of applying heavy elastomer coatings all the way to the tip of pipettes is presented. Noise sources arising from the pipette and its holder are described theoretically, and it is shown that measured noise is in good agreement with such predictions. With low noise capacitive feedback electronics, small geometry holders, and thick-walled quartz pipettes coated with low-loss elastomers we have been routinely able to achieve noise of 100 fA rms or less in a 5-kHz bandwidth with real cell patches and a pipette immersion depth of approximately 2 mm. On occasion we have achieved noise as low as 60 fA rms in this bandwidth.

Citing Articles

Nanopipettes as a Potential Diagnostic Tool for Selective Nanopore Detection of Biomolecules.

Kuanaeva R, Vaneev A, Gorelkin P, Erofeev A Biosensors (Basel). 2024; 14(12).

PMID: 39727892 PMC: 11674911. DOI: 10.3390/bios14120627.


A deep learning approach to real-time Markov modeling of ion channel gating.

Oikonomou E, Juli Y, Kolan R, Kern L, Gruber T, Alzheimer C Commun Chem. 2024; 7(1):280.

PMID: 39616256 PMC: 11608338. DOI: 10.1038/s42004-024-01369-y.


Direct whole-cell patch-clamp recordings from small boutons in rodent primary neocortical neuron cultures.

Ritzau-Jost A, Nerlich J, Kaas T, Krueger M, Tsintsadze T, Eilers J STAR Protoc. 2023; 4(2):102168.

PMID: 36920913 PMC: 10026040. DOI: 10.1016/j.xpro.2023.102168.


2D-dwell-time analysis with simulations of ion-channel gating using high-performance computing.

Oikonomou E, Gruber T, Chandra A, Holler S, Alzheimer C, Wellein G Biophys J. 2023; 122(7):1287-1300.

PMID: 36814379 PMC: 10111359. DOI: 10.1016/j.bpj.2023.02.023.


Small Size of Recorded Neuronal Structures Confines the Accuracy in Direct Axonal Voltage Measurements.

Olah V, Tarcsay G, Brunner J eNeuro. 2021; 8(4).

PMID: 34257077 PMC: 8342265. DOI: 10.1523/ENEURO.0059-21.2021.


References
1.
Hamill O, Marty A, Neher E, Sakmann B, Sigworth F . Improved patch-clamp techniques for high-resolution current recording from cells and cell-free membrane patches. Pflugers Arch. 1981; 391(2):85-100. DOI: 10.1007/BF00656997. View

2.
Rae J, Levis R . Glass technology for patch clamp electrodes. Methods Enzymol. 1992; 207:66-92. DOI: 10.1016/0076-6879(92)07005-9. View

3.
Levis R, Rae J . Constructing a patch clamp setup. Methods Enzymol. 1992; 207:14-66. DOI: 10.1016/0076-6879(92)07004-8. View

4.
Rae J, Levis R . Patch Clamp Recordings from the Epithelium of the Lens Obtained using Glasses Selected for Low Noise and Improved Sealing Properties. Biophys J. 2009; 45(1):144-6. PMC: 1435259. DOI: 10.1016/S0006-3495(84)84142-9. View