Quantitative Potentials of Dynamic Emission Computed Tomography

Overview
Journal J Nucl Med
Specialty Nuclear Medicine
Date 1978 Mar 1
PMID 632910
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Statistical uncertainties in emission computed tomography were simulated in 60 computer studies involving various numbers of events and distributions of activity. Previous studies have shown that for a uniform disc of activity of rms percentage of uncertainty per resolution cell is: 120 X (number of resolution cells)1/4 X (number of events per resolution cell)- 1/2. In this work we examined the more general situation where one or two regions of uniform activity are surrounded by a uniform background, and found that for an equal number of recorded events the uncertainties were reduced when the activity was concentrated in a portion of the field. The empirical relation rms % uncertainty in nt = 120(N)1/4(nt)-3/4, where nt is the number of events in an average target (organ) resolution cell and N is the total number of events recorded, satisfactorily described the relationships between uncertainties, contrast, total number of detected events, and number of resolution cells for all 60 computer studies. By means of this relation, we show the theoretical possibility of gated cardiac imaging with 20% uncertainty in 1 cm X 1 cm regions, and of 1-sec cerebral blood-flow images with 20% uncertainty in 2 cm X 2 cm regions.

Citing Articles

EANM practice guideline for quantitative SPECT-CT.

Dickson J, Armstrong I, Minguez Gabina P, Denis-Bacelar A, Krizsan A, Gear J Eur J Nucl Med Mol Imaging. 2022; 50(4):980-995.

PMID: 36469107 PMC: 9931838. DOI: 10.1007/s00259-022-06028-9.


System resolution versus image uncertainty for positron emission tomography scanners.

Studen A, Clinthorne N J Med Imaging (Bellingham). 2022; 9(3):033501.

PMID: 35578602 PMC: 9099252. DOI: 10.1117/1.JMI.9.3.033501.


A method to assess image quality for Low-dose PET: analysis of SNR, CNR, bias and image noise.

Yan J, Schaefferkoette J, Conti M, Townsend D Cancer Imaging. 2016; 16(1):26.

PMID: 27565136 PMC: 5002150. DOI: 10.1186/s40644-016-0086-0.


Properties of noise in positron emission tomography images reconstructed with filtered-backprojection and row-action maximum likelihood algorithm.

Teymurazyan A, Riauka T, Jans H, Robinson D J Digit Imaging. 2012; 26(3):447-56.

PMID: 22850933 PMC: 3649042. DOI: 10.1007/s10278-012-9511-5.


Selection of weighting factors for quantification of PET radioligand binding using simplified reference tissue models with noisy input functions.

Normandin M, Koeppe R, Morris E Phys Med Biol. 2012; 57(3):609-29.

PMID: 22241524 PMC: 3361066. DOI: 10.1088/0031-9155/57/3/609.