Doppler Hemodynamic Evaluation of Prosthetic (Starr-Edwards and Björk-Shiley) and Bioprosthetic (Hancock and Carpentier-Edwards) Cardiac Valves

Overview
Journal Am J Cardiol
Date 1985 Aug 1
PMID 4025173
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

One hundred thirty-four patients with prosthetic or bioprosthetic heart valves were investigated with Doppler echocardiography to determine normal values for commonly used prosthetic valves and to test the specificity of abnormal Doppler findings. In 70 patients the aortic valves had been replaced and in 64 the mitral valves had been replaced. Gradients across prostheses in the aortic position were calculated from maximal velocity. Peak calculated aortic transvalvular gradients in normal subjects were 22 +/- 10 mm Hg in 33 Björk-Shiley valves, 23 +/- 10 mm Hg in 27 porcine valves and 29 +/- 13 mm Hg in 6 Starr-Edwards valves. Mild aortic regurgitation was seen in 42% of Björk-Shiley valves, 26% of porcine valves and 2 of 6 Starr-Edwards valves. Mitral valve orifice was calculated by the pressure half-time method. In clinically normal patients with mitral valve prostheses, the effective mitral valve orifice was 2.5 +/- 0.8 cm2 in 35 Björk-Shiley valves, 2.1 +/- 0.7 cm2 in 17 porcine valves, and 2.0 +/- 0.3 cm2 in 10 Starr-Edwards valves. Mitral regurgitation was found in 11% of Björk-Shiley valves, 19% of porcine valves and 30% of Starr-Edwards valves. Repeat studies at 2 weeks to 14 months revealed no difference in 8 aortic and 14 mitral prostheses. Seven aortic and 4 mitral valves functioned abnormally as determined by Doppler, and the abnormal function was confirmed in each at surgery or at cardiac catheterization.(ABSTRACT TRUNCATED AT 250 WORDS)

Citing Articles

Left ventricular dynamics after aortic valve replacement: a long-term, combined radionuclide angiographic and ultrasonographic study.

Masotti C, Bonfranceschi P, Rusticali G, Rusticali F, Pierangeli A Tex Heart Inst J. 1992; 19(2):97-106.

PMID: 15227421 PMC: 326262.


Echocardiographic assessment of artificial heart valves: British Society of Echocardiography position paper.

Chambers J, Fraser A, Lawford P, Nihoyannopoulos P, Simpson I Br Heart J. 1994; 71(4 Suppl):6-14.

PMID: 8011398 PMC: 483704. DOI: 10.1136/hrt.71.4_suppl.6.


Non-invasive assessment by Doppler ultrasound of 155 patients with bioprosthetic valves: a comparison of the Wessex porcine, low profile Ionescu-Shiley, and Hancock pericardial bioprostheses.

Simpson I, Reece I, Houston A, Hutton I, Wheatley D, Cobbe S Br Heart J. 1986; 56(1):83-8.

PMID: 3524634 PMC: 1277389. DOI: 10.1136/hrt.56.1.83.


Oesophageal echocardiography.

Gussenhoven E, Taams M, Roelandt J, Bom K, Honkoop J, de Jong N Int J Card Imaging. 1987; 2(4):231-9.

PMID: 3323333 DOI: 10.1007/BF01784779.


Doppler measurement of cardiac output across prosthetic mitral valves.

Dittmann H, Voelker W, Karsch K, Seipel L Klin Wochenschr. 1990; 68(5):263-8.

PMID: 2325355 DOI: 10.1007/BF02116054.