» Articles » PMID: 4015919

Evaluation of an Elliptical Area Technique for Calculating Mitral Blood Flow by Doppler Echocardiography

Overview
Journal Br Heart J
Date 1985 Jul 1
PMID 4015919
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

To evaluate a method for measuring blood flow through the mitral valve 18 normal subjects and 19 patients with cardiac disease in whom mitral and aortic blood flows were identical were studied. Initially the mitral ring area was planimetered from the echocardiographic image, but the results of area calculation using the mathematical formula for the area of an ellipse were found to approximate to within 8% of the planimetered result in most cases. The formula was therefore used if the ring appeared elliptical on the cross sectional echo image, and other shapes were planimetered. Mitral velocity, aligned with flow in three planes, was recorded just distal to the ring. Mitral flow calculated using the elliptical technique correlated closely with flow measured in the ascending aorta by the Doppler technique and also with systemic flow measured by the Fick method at cardiac catheterisation in 10 patients. The mitral flow technique that assumed a circular orifice correlated almost as well with Doppler aortic flow and with Fick flow but overestimated flow by a mean of 1446 ml, whereas the elliptical method had a mean error of only 138 ml. Both methods correlated well with standards, but the elliptical method was easy to apply and gave a better correlation with comparison reference values.

Citing Articles

Quantitative Doppler-echocardiographic determination of regurgitant volume in patients with aortic insufficiency.

Schoenhagen P, Drude L, Klein H, Garcia M Open Cardiovasc Med J. 2009; 2:12-9.

PMID: 19590613 PMC: 2707760. DOI: 10.2174/1874192400802010012.


Automated quantification of aortic regurgitant volume and regurgitant fraction using the digital colour Doppler velocity profile integration method in patients with aortic regurgitation.

Miyake Y, Hozumi T, Mori I, Sugioka K, Yamamuro A, Akasaka T Heart. 2002; 88(5):481-4.

PMID: 12381638 PMC: 1767402. DOI: 10.1136/heart.88.5.481.


Magnetic resonance velocity mapping of normal human transmitral velocity profiles.

Fujimoto S, Mohiaddin R, Parker K, Gibson D Heart Vessels. 1995; 10(5):236-40.

PMID: 8903998 DOI: 10.1007/BF01744902.


Quantitative applications of Doppler cardiography in congenital heart disease.

Colan S Cardiovasc Intervent Radiol. 1987; 10(6):332-47.

PMID: 3123059 DOI: 10.1007/BF02577344.


Impact of changes in heart rate and stroke volume on the cross sectional flow velocity distribution of diastolic mitral blood flow. A study on 6 patients with pacemakers programmed at different heart rates.

Samstad S, Bathen J, Rossvoll O, Torp H, Skjaerpe T, Hatle L Int J Card Imaging. 1992; 8(2):75-83.

PMID: 1629642 DOI: 10.1007/BF01137528.


References
1.
Magnin P, Stewart J, Myers S, von Ramm O, Kisslo J . Combined doppler and phased-array echocardiographic estimation of cardiac output. Circulation. 1981; 63(2):388-92. DOI: 10.1161/01.cir.63.2.388. View

2.
Ormiston J, Shah P, Tei C, Wong M . Size and motion of the mitral valve annulus in man. I. A two-dimensional echocardiographic method and findings in normal subjects. Circulation. 1981; 64(1):113-20. DOI: 10.1161/01.cir.64.1.113. View

3.
Alverson D, Eldridge M, Dillon T, Yabek S, Berman Jr W . Noninvasive pulsed Doppler determination of cardiac output in neonates and children. J Pediatr. 1982; 101(1):46-50. DOI: 10.1016/s0022-3476(82)80178-9. View

4.
Goldberg S, Sahn D, ALLEN H, Hoenecke H, Carnahan Y . Evaluation of pulmonary and systemic blood flow by 2-dimensional Doppler echocardiography using fast Fourier transform spectral analysis. Am J Cardiol. 1982; 50(6):1394-400. DOI: 10.1016/0002-9149(82)90481-7. View

5.
Huntsman L, STEWART D, Barnes S, Franklin S, Colocousis J, Hessel E . Noninvasive Doppler determination of cardiac output in man. Clinical validation. Circulation. 1983; 67(3):593-602. DOI: 10.1161/01.cir.67.3.593. View