Abstract

The accumulated dose from sequential treatments of metachronous non-melanoma skin cancer can be assessed using image registration, although guidelines for selecting the appropriate algorithm are lacking. This study shows the impact of rigid (RIR), deformable (DIR) and deformable structure-based (SDIR) algorithms on the skin dose. DIR increased: the maximum dose (39.2 Gy vs 9.4 Gy), the dose to 0.1 cm (16.4 Gy vs 7.8 Gy) and the dose to 2 cm (7.6 Gy vs 5.7 Gy). RIR only affected the maximum dose, which increased to 17.0 Gy. SDIR correctly translated the dose maps, as none of the parameters changed significantly.

References
1.
Swamidas J, Kirisits C, De Brabandere M, Hellebust T, Siebert F, Tanderup K . Image registration, contour propagation and dose accumulation of external beam and brachytherapy in gynecological radiotherapy. Radiother Oncol. 2019; 143:1-11. DOI: 10.1016/j.radonc.2019.08.023. View

2.
Jamema S, Mahantshetty U, Andersen E, Noe K, Sorensen T, Kallehauge J . Uncertainties of deformable image registration for dose accumulation of high-dose regions in bladder and rectum in locally advanced cervical cancer. Brachytherapy. 2015; 14(6):953-62. DOI: 10.1016/j.brachy.2015.08.011. View

3.
Ouhib Z, Kasper M, Calatayud J, Rodriguez S, Bhatnagar A, Pai S . Aspects of dosimetry and clinical practice of skin brachytherapy: The American Brachytherapy Society working group report. Brachytherapy. 2015; 14(6):840-58. DOI: 10.1016/j.brachy.2015.06.005. View

4.
Szlag M, Kellas-Sleczka S, Wojcieszek P, Stankiewicz M, Cholewka A, Pruefer A . Advanced dose calculation algorithm in superficial brachytherapy - the impact of tissue inhomogeneity on treatment plan dosimetry. J Contemp Brachytherapy. 2021; 13(4):441-446. PMC: 8407266. DOI: 10.5114/jcb.2021.106541. View

5.
Gonzalez-Perez V, Rembielak A, Guinot J, Jaberi R, Lancellotta V, Walter R . H&N and Skin (HNS) GEC-ESTRO Working Group critical review of recommendations regarding prescription depth, bolus thickness and maximum dose in skin superficial brachytherapy with flaps and customized moulds. Radiother Oncol. 2022; 175:122-132. DOI: 10.1016/j.radonc.2022.08.022. View