Precision and Accuracy of Four Handheld Blood Lactate Analyzers Across Low to High Exercise Intensities

Overview
Specialty Physiology
Date 2024 Aug 19
PMID 39158591
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To evaluate the precision and accuracy in measured blood lactate concentrations among four commonly used handheld lactate analyzers compared to two stationary analyzers.

Methods: Venous blood samples were taken at exercise intensities ranging from low to high. The blood lactate concentration was measured simultaneously with four pairs of handheld lactate analyzers (two new units of each brand: Lactate Plus, Lactate Pro2, Lactate Scout 4, and TaiDoc TD-4289), and compared with two stationary analyzers (Biosen C-Line and YSI Sport 1500). Measurements were repeated for a range of blood lactate concentrations (measured with Biosen) from 0.88 to 4.89 mM with a median difference between measurements of 0.10 mM.

Results: The mean relative differences to the Biosen analyzer were 7% (Plus), 7% (Pro), 10% (Scout), 42% (Tai), and 32% (Ysi). The residual standard errors after linear regression against Biosen were 0.18 mM (Plus), 0.20 mM (Pro), 0.22 mM (Scout), 0.15 mM (Tai), and 0.06 mM (Ysi). Accordingly, a blood lactate concentration of 3 mM measured with Biosen yielded 95% prediction intervals that were 0.72 mM (Plus), 0.80 mM (Pro), 0.87 mM (Scout), 0.60 mM (Tai), and 0.23 mM (Ysi) wide.

Conclusion: Compared to our two stationary analyzers, the precision of the four handheld lactate analyzers evaluated in this study was poor. Among the four, Tai was the most precise; however, this analyzer had low accuracy with a substantial mean difference to the reference analyzer.

References
1.
Bagger M, PETERSEN P, Pedersen P . Biological variation in variables associated with exercise training. Int J Sports Med. 2003; 24(6):433-40. DOI: 10.1055/s-2003-41180. View

2.
Bonaventura J, Sharpe K, Knight E, Fuller K, Tanner R, Gore C . Reliability and accuracy of six hand-held blood lactate analysers. J Sports Sci Med. 2015; 14(1):203-14. PMC: 4306774. View

3.
Bourdon P, Woolford S, Buckley J . Effects of Varying the Step Duration on the Determination of Lactate Thresholds in Elite Rowers. Int J Sports Physiol Perform. 2017; 13(6):687-693. DOI: 10.1123/ijspp.2017-0258. View

4.
Calatayud O, Tenias J . Effects of time, temperature and blood cell counts on levels of lactate in heparinized whole blood gas samples. Scand J Clin Lab Invest. 2003; 63(4):311-4. View

5.
Casado A, Foster C, Bakken M, Tjelta L . Does Lactate-Guided Threshold Interval Training within a High-Volume Low-Intensity Approach Represent the "Next Step" in the Evolution of Distance Running Training?. Int J Environ Res Public Health. 2023; 20(5). PMC: 10000870. DOI: 10.3390/ijerph20053782. View