Deep Learning-aided Respiratory Motion Compensation in PET/CT: Addressing Motion Induced Resolution Loss, Attenuation Correction Artifacts and PET-CT Misalignment

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Respiratory motion (RM) significantly impacts image quality in thoracoabdominal PET/CT imaging. This study introduces a unified data-driven respiratory motion correction (uRMC) method, utilizing deep learning neural networks, to solve all the major issues caused by RM, i.e., PET resolution loss, attenuation correction artifacts, and PET-CT misalignment.

Methods: In a retrospective study, 737 patients underwent [F]FDG PET/CT scans using the uMI Panorama PET/CT scanner. Ninety-nine patients, who also had respiration monitoring device (VSM), formed the validation set. The remaining data of the 638 patients were used to train neural networks used in the uRMC. The uRMC primarily consists of three key components: (1) data-driven respiratory signal extraction, (2) attenuation map generation, and (3) PET-CT alignment. SUV metrics were calculated within 906 lesions for three approaches, i.e., data-driven uRMC (proposed), VSM-based uRMC, and OSEM without motion correction (NMC). RM magnitude of major organs were estimated.

Results: uRMC enhanced diagnostic capabilities by revealing previously undetected lesions, sharpening lesion contours, increasing SUV values, and improving PET-CT alignment. Compared to NMC, uRMC showed increases of 10% and 17% in SUV and SUV across 906 lesions. Sub-group analysis showed significant SUV increases in small and medium-sized lesions with uRMC. Minor differences were found between VSM-based and data-driven uRMC methods, with the SUV was found statistically marginal significant or insignificant between the two methods. The study observed varied motion amplitudes in major organs, typically ranging from 10 to 20 mm.

Conclusion: A data-driven solution for respiratory motion in PET/CT has been developed, validated and evaluated. To the best of our knowledge, this is the first unified solution that compensates for the motion blur within PET, the attenuation mismatch artifacts caused by PET-CT misalignment, and the misalignment between PET and CT.

Citing Articles

Quantitative analysis of patient motion in walk-through PET scanner and standard axial field of view pet scanner using infrared-based tracking.

Aziz R, Maebe J, Muller F, DAsseler Y, Vandenberghe S EJNMMI Phys. 2024; 11(1):99.

PMID: 39581945 PMC: 11586328. DOI: 10.1186/s40658-024-00704-5.

References
1.
Ren S, Jin X, Chan C, Jian Y, Mulnix T, Liu C . Data-driven event-by-event respiratory motion correction using TOF PET list-mode centroid of distribution. Phys Med Biol. 2017; 62(12):4741-4755. PMC: 6048592. DOI: 10.1088/1361-6560/aa700c. View

2.
Allen-Auerbach M, Yeom K, Park J, Phelps M, Czernin J . Standard PET/CT of the chest during shallow breathing is inadequate for comprehensive staging of lung cancer. J Nucl Med. 2006; 47(2):298-301. View

3.
Hwang D, Kang S, Kim K, Choi H, Seo S, Lee J . Data-driven respiratory phase-matched PET attenuation correction without CT. Phys Med Biol. 2021; 66(11). DOI: 10.1088/1361-6560/abfc8f. View

4.
Hashimoto F, Onishi Y, Ote K, Tashima H, Reader A, Yamaya T . Deep learning-based PET image denoising and reconstruction: a review. Radiol Phys Technol. 2024; 17(1):24-46. PMC: 10902118. DOI: 10.1007/s12194-024-00780-3. View

5.
Chan C, Onofrey J, Jian Y, Germino M, Papademetris X, Carson R . Non-Rigid Event-by-Event Continuous Respiratory Motion Compensated List-Mode Reconstruction for PET. IEEE Trans Med Imaging. 2017; 37(2):504-515. PMC: 7304524. DOI: 10.1109/TMI.2017.2761756. View