Puncture Cube Patient-Mounted Navigation System Versus Freehand Method for CT-Guided Needle Placement: Study on a Neoprene Covered Elliptical Cylinder Gelatin Phantom

Overview
Date 2024 Jul 10
PMID 38987424
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The study aims to show how the "Puncture Cube" (PC) (Medical Templates, Egg, Switzerland) compares to the freehand method (FHM) for CT-guided punctures.

Methods: The PC is a patient-mounted disassemblable cube consisting of an upper and lower template with multiple holes each to predefine puncture trajectory. A total of 80 punctures (FHM in-plane, FHM off-plane, PC in-plane, PC off-plane) was performed by 4 radiologists on a target 9.1 cm below surface level of a neoprene covered elliptical cylinder gelatin phantom. The PC was never disassembled. Evaluated parameters were procedure time, number of CT-scans, euclidean distance (ED) and normal distance (ND). Respective parameters of FHM and PC were compared using the Wilcoxon signed-rank test and Levene test with significance levels of 5%.

Results: PC achieved smaller ED and ND values after initial needle insertion without corrections for both in-plane and off-plane punctures (P > 0.05). Variance of initial NDs was off-plane significantly larger for FHM. Final ED after needle path corrections was smaller for FHM both in- and off-plane (P < 0.05). Final off-plane ND was significantly lower for FHM with no significant difference in final in-plane ND. FHM off-plane punctures were significantly faster. There was no significant difference in CT-scans between both methods.

Conclusion: Utilizing the PC may improve initial needle positioning and safety especially off-plane. However, better final needle positioning after correction with the greater freedom of movement method may suggest need for disassembly of the cube.

References
1.
Marel M, Padr R, Fila L, Rakita D, Mendez F, Capkova L . Biopsy of lung lesions under CT control. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2020; 165(4):390-394. DOI: 10.5507/bp.2020.040. View

2.
Wessendorf J, Konig A, Heers H, Mahnken A . Repeat Percutaneous Radiofrequency Ablation of T1 Renal Cell Carcinomas is Safe in Patients with Von Hippel-Lindau Disease. Cardiovasc Intervent Radiol. 2021; 44(12):2022-2025. PMC: 8626382. DOI: 10.1007/s00270-021-02935-w. View

3.
Nishikawa H, Inuzuka T, Takeda H, Nakajima J, Matsuda F, Sakamoto A . Comparison of percutaneous radiofrequency thermal ablation and surgical resection for small hepatocellular carcinoma. BMC Gastroenterol. 2011; 11:143. PMC: 3260104. DOI: 10.1186/1471-230X-11-143. View

4.
Yang Y, Yu H, Tan X, You Y, Liu F, Zhao T . Liver resection versus radiofrequency ablation for recurrent hepatocellular carcinoma: a systematic review and meta-analysis. Int J Hyperthermia. 2021; 38(1):875-886. DOI: 10.1080/02656736.2021.1933218. View

5.
Vogl T, Nour-Eldin N, Hammerstingl R, Panahi B, Naguib N . Microwave Ablation (MWA): Basics, Technique and Results in Primary and Metastatic Liver Neoplasms - Review Article. Rofo. 2017; 189(11):1055-1066. DOI: 10.1055/s-0043-117410. View