Comparison of Intraocular Pressure Measurements with the Tono-pen, Goldmann Applanation Tonometer, and Noncontact Tonometer in Nonglaucomatous Pseudophakic Children

Overview
Journal Int Ophthalmol
Specialty Ophthalmology
Date 2024 Jun 27
PMID 38935310
Authors
Affiliations
Soon will be listed here.
Abstract

Background: To evaluate the agreement between the Goldman applanation tonometer (GAT), Tono-Pen, and noncontact tonometer (NCT) in the measurement of intraocular pressure (IOP) in pseudophakic children.

Methods: The medical records of nonglaucomatous pseudophakic children between 2009 and 2019 were retrospectively analyzed. A total of 46 eyes of 23 patients operated for bilateral pediatric cataract were included in the study. The patients' mean age was 13.4 ± 4.1 years. Central corneal thickness (CCT) and IOP values measured with the GAT, Tono-Pen, and NCT were recorded. Agreement between the tonometers was evaluated by intraclass correlation coefficients (ICC) and the Bland-Altman method.

Results: The mean IOP of the 46 eyes included in the study was measured as 13.7 ± 2.3 mm Hg with the GAT, 16.0 ± 2.3 mm Hg with NCT, and 16.5 ± 2.3 mm Hg with the Tono-Pen (p < 0.001). There is no statistical difference between NCT and Tono-Pen measurements, while GAT measurements were significantly lower than those of the NCT and Tono-pen. ICC values showed fair agreement between NCT and Tono-Pen (ICC = 0.720), whereas there was poor agreement between GAT and NCT (ICC = 0.501) and Tono-pen (ICC = 0.314).

Conclusions: With all devices included in the study, thicker corneas were associated with higher IOP measurements. Although there was moderate agreement between the NCT and Tono-Pen, there was a statistically significant difference in the IOP values provided by the three devices. Our results suggest these devices should not be used interchangeably.

References
1.
Wu X, Long E, Lin H, Liu Y . Prevalence and epidemiological characteristics of congenital cataract: a systematic review and meta-analysis. Sci Rep. 2016; 6:28564. PMC: 4917826. DOI: 10.1038/srep28564. View

2.
Sorsby A, Sheridan M . The eye at birth: measurement of the principal diameters in forty-eight cadavers. J Anat. 1960; 94:192-7. PMC: 1244392. View

3.
Sheeladevi S, Lawrenson J, Fielder A, Suttle C . Global prevalence of childhood cataract: a systematic review. Eye (Lond). 2016; 30(9):1160-9. PMC: 5023808. DOI: 10.1038/eye.2016.156. View

4.
Crockett C, Camero K, Kong L, Yen K . Visual outcomes of patients presenting with bilateral infantile cataracts and nystagmus. Can J Ophthalmol. 2017; 52(2):203-206. DOI: 10.1016/j.jcjo.2016.08.021. View

5.
Brandt J . Corneal thickness in glaucoma screening, diagnosis, and management. Curr Opin Ophthalmol. 2004; 15(2):85-9. DOI: 10.1097/00055735-200404000-00004. View