Abstract

Background: Closed-loop oxygen control systems automatically adjust the fraction of inspired oxygen (FiO) to maintain oxygen saturation (SpO) within a predetermined target range. Their performance with low and high-flow oxygen therapies, but not with non-invasive ventilation, has been established. We compared the effect of automated oxygen on achieving and maintaining a target SpO range with nasal high flow (NHF), bilevel positive airway pressure (bilevel) and continuous positive airway pressure (CPAP), in stable hypoxaemic patients with chronic cardiorespiratory disease.

Methods: In this open-label, three-way cross-over trial, participants with resting hypoxaemia (n=12) received each of NHF, bilevel and CPAP treatments, in random order, with automated oxygen titrated for 10 min, followed by 36 min of standardised manual oxygen adjustments. The primary outcome was the time taken to reach target SpO range (92%-96%). Secondary outcomes included time spent within target range and physiological responses to automated and manual oxygen adjustments.

Results: Two participants were randomised to each of six possible treatment orders. During automated oxygen control (n=12), the mean (±SD) time to reach target range was 114.8 (±87.9), 56.6 (±47.7) and 67.3 (±61) seconds for NHF, bilevel and CPAP, respectively, mean difference 58.3 (95% CI 25.0 to 91.5; p=0.002) and 47.5 (95% CI 14.3 to 80.7; p=0.007) seconds for bilevel and CPAP versus NHF, respectively. Proportions of time spent within target range were 68.5% (±16.3), 65.6% (±28.7) and 74.7% (±22.6) for NHF, bilevel and CPAP, respectively.Manually increasing, then decreasing, the FiO resulted in similar increases and then decreases in SpO and transcutaneous carbon dioxide (PtCO) with NHF, bilevel and CPAP.

Conclusion: The target SpO range was achieved more quickly when automated oxygen control was initiated with bilevel and CPAP compared with NHF while time spent within the range across the three therapies was similar. Manually changing the FiO had similar effects on SpO and PtCO across each of the three therapies.

Trial Registration Number: ACTRN12622000433707.

References
1.
Trottier M, Bouchard P, LHer E, Lellouche F . Automated Oxygen Titration During CPAP and Noninvasive Ventilation in Healthy Subjects With Induced Hypoxemia. Respir Care. 2023; 68(11):1553-1560. PMC: 10589107. DOI: 10.4187/respcare.09866. View

2.
Crossley D, McGuire G, Barrow P, Houston P . Influence of inspired oxygen concentration on deadspace, respiratory drive, and PaCO2 in intubated patients with chronic obstructive pulmonary disease. Crit Care Med. 1997; 25(9):1522-6. DOI: 10.1097/00003246-199709000-00019. View

3.
Hansen E, Hove J, Bech C, Jensen J, Kallemose T, Vestbo J . Automated oxygen control with O2matic during admission with exacerbation of COPD. Int J Chron Obstruct Pulmon Dis. 2018; 13:3997-4003. PMC: 6300382. DOI: 10.2147/COPD.S183762. View

4.
Lellouche F, Bouchard P, Roberge M, Simard S, LHer E, Maltais F . Automated oxygen titration and weaning with FreeO2 in patients with acute exacerbation of COPD: a pilot randomized trial. Int J Chron Obstruct Pulmon Dis. 2016; 11:1983-90. PMC: 5003517. DOI: 10.2147/COPD.S112820. View

5.
Groves N, Tobin A . High flow nasal oxygen generates positive airway pressure in adult volunteers. Aust Crit Care. 2007; 20(4):126-31. DOI: 10.1016/j.aucc.2007.08.001. View