Abstract

Introduction: Hypoxaemia and hyperoxaemia may occur after surgery, with related complications. This multicentre randomised trial evaluated the impact of automated closed-loop oxygen administration after high-risk abdominal or thoracic surgeries in terms of optimising the oxygen saturation measured by pulse oximetry time within target range.

Methods: After extubation, patients with an intermediate to high risk of post-operative pulmonary complications were randomised to "standard" or "automated" closed-loop oxygen administration. The primary outcome was the percentage of time within the oxygenation range, during a 3-day frame. The secondary outcomes were the time with hypoxaemia and hyperoxaemia under oxygen.

Results: Among the 200 patients, time within range was higher in the automated group, both initially (≤3 h; 91.4±13.7% 40.2±35.1% of time, difference +51.0% (95% CI -42.8-59.2%); p<0.0001) and during the 3-day period (94.0±11.3% 62.1±23.3% of time, difference +31.9% (95% CI 26.3-37.4%); p<0.0001). Periods of hypoxaemia were reduced in the automated group (≤3 days; 32.6±57.8 min (1.2±1.9%) 370.5±594.3 min (5.0±11.2%), difference -10.2% (95% CI -13.9--6.6%); p<0.0001), as well as hyperoxaemia under oxygen (≤3 days; 5.1±10.9 min (4.8±11.2%) 177.9±277.2 min (27.0±23.8%), difference -22.0% (95% CI -27.6--16.4%); p<0.0001). Kaplan-Meier analysis depicted a significant difference in terms of hypoxaemia (p=0.01) and severe hypoxaemia (p=0.0003) occurrence between groups in favour of the automated group. 25 patients experienced hypoxaemia for >10% of the entire monitoring time during the 3 days within the standard group, as compared to the automated group (p<0.0001).

Conclusion: Automated closed-loop oxygen administration promotes greater time within the oxygenation target, as compared to standard manual administration, thus reducing the occurrence of hypoxaemia and hyperoxaemia.

Citing Articles

High-velocity nasal insufflation with oxygen assist module versus conventional high-velocity nasal insufflation after extubation: an open-label randomized crossover study.

Visuthisakchai S, Lersritwimanmaen P, Rittayamai N J Thorac Dis. 2025; 17(1):333-343.

PMID: 39975739 PMC: 11833560. DOI: 10.21037/jtd-24-1345.


Closed-loop oxygen usage during invasive mechanical ventilation of pediatric patients (CLOUDIMPP): a randomized controlled cross-over study.

Atakul G, Ceylan G, Sandal O, Soydan E, Hepduman P, Colak M Front Med (Lausanne). 2024; 11:1426969.

PMID: 39318593 PMC: 11420134. DOI: 10.3389/fmed.2024.1426969.


Automated Oxygen Administration in Hospitals: Will It Replace Manual Systems in the Near Future?.

ODriscoll B, Kirton L Respir Care. 2024; 69(9):1214-1217.

PMID: 39181719 PMC: 11349593. DOI: 10.4187/respcare.12230.


Effect of automated titration of oxygen on time spent in a prescribed oxygen saturation range in adults in the ICU after cardiac surgery.

Kirton L, Cruz R, Navarra L, Eathorne A, Cook J, Beasley R Crit Care Resusc. 2024; 26(2):64-70.

PMID: 39072230 PMC: 11282340. DOI: 10.1016/j.ccrj.2024.01.001.


Automated oxygen titration with non-invasive ventilation in hypoxaemic adults with cardiorespiratory disease: a randomised cross-over trial.

Kirton L, Kung S, Bird G, Black M, Semprini R, Eathorne A BMJ Open Respir Res. 2024; 11(1).

PMID: 38897612 PMC: 11191803. DOI: 10.1136/bmjresp-2023-002196.