Evaluation of a Novel Cuffless Photoplethysmography-based Wristband for Measuring Blood Pressure According to the Regulatory Standards

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

Aims: Elevated blood pressure (BP) is a key risk factor in cardiovascular diseases. However, obtaining reliable and reproducible BP remains a challenge. This study, therefore, aimed to evaluate a novel cuffless wristband, based on photoplethysmography (PPG), for continuous BP monitoring.

Methods And Results: Predictions by a PPG-guided algorithm were compared to arterial BP measurements (in the sub-clavian artery), obtained during cardiac catheterization. Eligible patients were included and screened based on AAMI/European Society of Hypertension (ESH)/ISO Universal Standard requirements. The machine learning-based BP algorithm required three cuff-based initialization measurements in combination with ∼100 features (signal-derived and patient demographic-based). Ninety-seven patients and 420 samples were included. Mean age, weight, and height were 67.1 years (SD 11.1), 83.4 kg (SD 16.1), and 174 cm (SD 10), respectively. Systolic BP was ≤100 mmHg in 48 samples (11%) and ≥160 mmHg in 106 samples (25%). Diastolic BP was ≤70 mmHg in 222 samples (53%) and ≥85 mmHg in 99 samples (24%). The algorithm showed mean errors of ±3.7 mmHg (SD 4.4 mmHg) and ±2.5 mmHg (SD 3.7 mmHg) for systolic and diastolic BP, respectively. Similar results were observed across all genders and skin colours (Fitzpatrick I-VI).

Conclusion: This study provides initial evidence for the accuracy of a PPG-based BP algorithm in combination with a cuffless wristband across a range of BP distributions. This research complies with the AAMI/ESH/ISO Universal Standard, however, further research is required to evaluate the algorithms performance in light of the remaining European Society of Hypertension recommendations.

Clinical Trial Registration: www.clinicaltrials.gov, NCT05566886.

Citing Articles

AI for the hemodynamic assessment of critically ill and surgical patients: focus on clinical applications.

Michard F, Mulder M, Gonzalez F, Sanfilippo F Ann Intensive Care. 2025; 15(1):26.

PMID: 39992575 PMC: 11850697. DOI: 10.1186/s13613-025-01448-w.


Performance of a medical smartband with photoplethysmography technology and artificial intelligence algorithm to detect atrial fibrillation.

Blok S, Gielen W, Piek M, Hoeksema W, Tulevski I, Somsen G Mhealth. 2025; 11:5.

PMID: 39944860 PMC: 11811644. DOI: 10.21037/mhealth-24-10.


New sensors for the early detection of clinical deterioration on general wards and beyond - a clinician's perspective.

Michard F, Saugel B J Clin Monit Comput. 2024; .

PMID: 39546216 DOI: 10.1007/s10877-024-01235-1.


Clinical evaluation of a smart wristband for monitoring oxygen saturation, pulse rate, and respiratory rate.

Monnink S, van Vliet M, Kuiper M, Constandse J, Hoftijzer D, Muller M J Clin Monit Comput. 2024; .

PMID: 39388061 DOI: 10.1007/s10877-024-01229-z.

References
1.
Scheer B, Perel A, Pfeiffer U . Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Crit Care. 2002; 6(3):199-204. PMC: 137445. DOI: 10.1186/cc1489. View

2.
Cambiaso-Daniel J, Rontoyanni V, Foncerrada G, Nguyen A, Capek K, Wurzer P . Correlation between invasive and noninvasive blood pressure measurements in severely burned children. Burns. 2018; 44(7):1787-1791. PMC: 6387786. DOI: 10.1016/j.burns.2018.03.001. View

3.
Picone D, Schultz M, Otahal P, Aakhus S, Al-Jumaily A, Black J . Accuracy of Cuff-Measured Blood Pressure: Systematic Reviews and Meta-Analyses. J Am Coll Cardiol. 2017; 70(5):572-586. DOI: 10.1016/j.jacc.2017.05.064. View

4.
Islam S, Chow C, Daryabeygikhotbehsara R, Subedi N, Rawstorn J, Tegegne T . Wearable cuffless blood pressure monitoring devices: a systematic review and meta-analysis. Eur Heart J Digit Health. 2023; 3(2):323-337. PMC: 9708022. DOI: 10.1093/ehjdh/ztac021. View

5.
Bradley C, Shimbo D, Colburn D, Pugliese D, Padwal R, Sia S . Cuffless Blood Pressure Devices. Am J Hypertens. 2022; 35(5):380-387. PMC: 9088838. DOI: 10.1093/ajh/hpac017. View