Comparison Between Cuff-based and Invasive Systolic Blood Pressure Amplification

Abstract

Objective: Accurate measurement of central blood pressure (BP) using upper arm cuff-based methods is associated with several factors, including determining the level of systolic BP (SBP) amplification. This study aimed to determine the agreement between cuff-based and invasively measured SBP amplification.

Methods: Patients undergoing coronary angiography had invasive SBP amplification (brachial SBP - central SBP) measured simultaneously with cuff-based SBP amplification using a commercially available central BP device (device 1: Sphygmocor Xcel; n = 171, 70% men, 60 ± 10 years) and a now superseded model of a central BP device (device 2: Uscom BP+; n = 52, 83% men, 62 ± 10 years).

Results: Mean difference (±2SD, limits of agreement) between cuff-based and invasive SBP amplification was 4 mmHg (-12, +20 mmHg, P < 0.001) for device 1 and -2 mmHg (-14, +10 mmHg, P = 0.10) for device 2. Both devices systematically overestimated SBP amplification at lower levels and underestimated at higher levels of invasive SBP amplification, but with stronger bias for device 1 (r = -0.68 vs. r = -0.52; Z = 2.72; P = 0.008). Concordance of cuff-based and invasive SBP amplification across quartiles of invasive SBP amplification was low, particularly in the lowest and highest quartiles. The root mean square errors from regression between cuff-based central SBP and brachial SBP were significantly lower (indicating less variability) than from invasive regression models (P < 0.001).

Conclusions: Irrespective of the difference from invasive measurements, cuff-based estimates of SBP amplification showed evidence of proportional systematic bias and had less individual variability. These observations could provide insights on how to improve the performance of cuff-based central BP.

Citing Articles

Personalised hypertension management with accurate blood pressure measurement: much achieved, much more to do.

Sharman J Hypertens Res. 2025; .

PMID: 39972181 DOI: 10.1038/s41440-025-02155-6.


Central blood pressure and peripheral augmentation index following acute submaximal arm versus leg exercise.

Wu Z, Heffernan K Eur J Appl Physiol. 2024; 124(12):3729-3739.

PMID: 39133269 DOI: 10.1007/s00421-024-05573-5.


Evaluation of a novel cuffless photoplethysmography-based wristband for measuring blood pressure according to the regulatory standards.

van Vliet M, Monnink S, Kuiper M, Constandse J, Hoftijzer D, Ronner E Eur Heart J Digit Health. 2024; 5(3):335-343.

PMID: 38774367 PMC: 11104472. DOI: 10.1093/ehjdh/ztae006.


Central-to-peripheral blood pressure amplification: role of the recording site, technology, analysis approach, and calibration scheme in invasive and non-invasive data agreement.

Zocalo Y, Bia D, Sanchez R, Lev G, Mendiz O, Ramirez A Front Cardiovasc Med. 2023; 10:1256221.

PMID: 37886732 PMC: 10598655. DOI: 10.3389/fcvm.2023.1256221.


Accuracy of cuff blood pressure and systolic blood pressure amplification.

Bui T, Picone D, Schultz M, Peng X, Black J, Dwyer N Hypertens Res. 2023; 46(8):1961-1969.

PMID: 37217732 PMC: 10404511. DOI: 10.1038/s41440-023-01311-0.