» Articles » PMID: 38686202

Accuracy of Continuous Glucose Monitoring During Exercise-related Hypoglycemia in Individuals with Type 1 Diabetes

Overview
Specialty Endocrinology
Date 2024 Apr 30
PMID 38686202
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Hypoglycemia is common in individuals with type 1 diabetes, especially during exercise. We investigated the accuracy of two different continuous glucose monitoring systems during exercise-related hypoglycemia in an experimental setting.

Materials And Methods: Fifteen individuals with type 1 diabetes participated in two separate euglycemic-hypoglycemic clamp days (Clamp-exercise and Clamp-rest) including five phases: 1) baseline euglycemia, 2) plasma glucose (PG) decline ± exercise, 3) 15-minute hypoglycemia ± exercise, 4) 45-minute hypoglycemia, and 5) recovery euglycemia. Interstitial PG levels were measured every five minutes, using Dexcom G6 (DG6) and FreeStyle Libre 1 (FSL1). Yellow Springs Instruments 2900 was used as PG reference method, enabling mean absolute relative difference (MARD) assessment for each phase and Clarke error grid analysis for each day.

Results: Exercise had a negative effect on FSL1 accuracy in phase 2 and 3 compared to rest (ΔMARD = +5.3 percentage points [(95% CI): 1.6, 9.1] and +13.5 percentage points [6.4, 20.5], respectively). In contrast, exercise had a positive effect on DG6 accuracy during phase 2 and 4 compared to rest (ΔMARD = -6.2 percentage points [-11.2, -1.2] and -8.4 percentage points [-12.4, -4.3], respectively). Clarke error grid analysis showed a decrease in clinically acceptable treatment decisions during Clamp-exercise for FSL1 while a contrary increase was observed for DG6.

Conclusion: Physical exercise had clinically relevant impact on the accuracy of the investigated continuous glucose monitoring systems and their ability to accurately detect hypoglycemia.

References
1.
. Glucose Concentrations of Less Than 3.0 mmol/L (54 mg/dL) Should Be Reported in Clinical Trials: A Joint Position Statement of the American Diabetes Association and the European Association for the Study of Diabetes. Diabetes Care. 2016; 40(1):155-157. DOI: 10.2337/dc16-2215. View

2.
Olafsdottir A, Attvall S, Sandgren U, Dahlqvist S, Pivodic A, Skrtic S . A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes. Diabetes Technol Ther. 2017; 19(3):164-172. PMC: 5359691. DOI: 10.1089/dia.2016.0392. View

3.
Finn M, Sherlock M, Feehan S, Guinan E, Moore K . Adherence to physical activity recommendations and barriers to physical activity participation among adults with type 1 diabetes. Ir J Med Sci. 2021; 191(4):1639-1646. PMC: 9308574. DOI: 10.1007/s11845-021-02741-w. View

4.
Coates A, Cohen J, Burr J . Investigating sensor location on the effectiveness of continuous glucose monitoring during exercise in a non-diabetic population. Eur J Sport Sci. 2023; 23(10):2109-2117. DOI: 10.1080/17461391.2023.2174452. View

5.
Brazeau A, Rabasa-Lhoret R, Strychar I, Mircescu H . Barriers to physical activity among patients with type 1 diabetes. Diabetes Care. 2008; 31(11):2108-9. PMC: 2571055. DOI: 10.2337/dc08-0720. View