Dosimetric Validation of Intensity-modulated Bolus Electron Conformal Therapy Planning and Delivery Using an Anthropomorphic Cylindrical Head Phantom

Overview
Date 2024 Apr 5
PMID 38576174
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: This work investigated the dosimetric accuracy of the intensity-modulated bolus electron conformal therapy (IM-BECT) planning and delivery process using the decimal ElectronRT (eRT) treatment planning system.

Methods: An IM-BECT treatment plan was designed using eRT for a cylindrical, anthropomorphic retromolar trigone phantom. Treatment planning involved specification of beam parameters and design of a variable thickness wax bolus and Passive Radiotherapy Intensity Modulator for Electrons (PRIME) device, which was comprised of 33 tungsten island blocks of discrete diameters from 0.158 to 0.223 cm (Intensity Reduction Factors from 0.937 to 0.875, respectively) inside a 10.1 × 6.7 cm copper cutout. For comparison of calculation accuracy, a BECT plan was generated by copying the IM-BECT plan and removing the intensity modulation. For both plans, a 16 MeV electron beam was used with 104.7 cm source-to-surface distance to bolus. In-phantom TLD-100 measurements (N = 47) were compared with both eRT planned dose distributions, which used the pencil beam redefinition algorithm with modifications for passive electron intensity modulation (IM-PBRA). Dose difference and distance to agreement (DTA) metrics were computed for each measurement point.

Results: Comparison of measured dose distributions with planned dose distributions yielded dose differences (calculated minus measured) characterized by a mean and standard deviation of -0.36% ± 1.64% for the IM-BECT plan, which was similar to -0.36% ± 1.90% for the BECT plan. All dose measurements were within 5% of the planned dose distribution, with both the BECT and IM-BECT measurement sets having 46/47 (97.8%) points within 3% or within 3 mm of the respective treatment plans.

Conclusions: It was found that the IM-BECT treatment plan generated using eRT was sufficiently accurate for clinical use when compared to TLD measurements in a cylindrical, anthropomorphic phantom, and was similarly accurate to the BECT treatment plan in the same phantom.

Citing Articles

Dosimetric validation of intensity-modulated bolus electron conformal therapy planning and delivery using an anthropomorphic cylindrical head phantom.

McGuffey A, Pitcher G, Erhart K, Hogstrom K J Appl Clin Med Phys. 2024; 25(7):e14347.

PMID: 38576174 PMC: 11244665. DOI: 10.1002/acm2.14347.

References
1.
Lukowiak M, Jezierska K, Boehlke M, Wiecko M, Lukowiak A, Podraza W . Utilization of a 3D printer to fabricate boluses used for electron therapy of skin lesions of the eye canthi. J Appl Clin Med Phys. 2017; 18(1):76-81. PMC: 5689892. DOI: 10.1002/acm2.12013. View

2.
Kim M, Kudchadker R, Kanke J, Zhang S, Perkins G . Bolus electron conformal therapy for the treatment of recurrent inflammatory breast cancer: a case report. Med Dosim. 2011; 37(2):208-13. DOI: 10.1016/j.meddos.2011.07.004. View

3.
Schneider U, Pedroni E, Lomax A . The calibration of CT Hounsfield units for radiotherapy treatment planning. Phys Med Biol. 1996; 41(1):111-24. DOI: 10.1088/0031-9155/41/1/009. View

4.
Kudchadker R, Hogstrom K, Garden A, McNeese M, Boyd R, Antolak J . Electron conformal radiotherapy using bolus and intensity modulation. Int J Radiat Oncol Biol Phys. 2002; 53(4):1023-37. DOI: 10.1016/s0360-3016(02)02811-0. View

5.
Opp D, Forster K, Li W, Zhang G, Harris E . Evaluation of bolus electron conformal therapy compared with conventional techniques for the treatment of left chest wall postmastectomy in patients with breast cancer. Med Dosim. 2013; 38(4):448-53. DOI: 10.1016/j.meddos.2013.08.002. View