Investigation of the Performance of Various Low-Cost Radon Monitors Under Variable Environmental Conditions

Overview
Journal Sensors (Basel)
Publisher MDPI
Specialty Biotechnology
Date 2024 Mar 28
PMID 38544099
Authors
Affiliations
Soon will be listed here.
Abstract

A comparison of low-cost radon monitors was conducted at the Laboratory of Natural Radiation (LNR). The monitors we evaluated were EcoQube, RadonEye, RadonEye Plus2, Spirit, ViewPlus, ViewRadon and WavePlus. An AlphaGUARD monitor calibrated at the Laboratory of Environmental Radioactivity of the University of Cantabria (LaRUC), accredited for testing and calibration according to ISO/IEC 17025, provided the reference value of radon concentration. The temporal stability of the monitors was studied, obtaining a percentage of missing records ranged from 1% to 19% of the data. The main technical characteristics studied were temporal stability, measurement ranges, accuracy, correlation and response time. The main results show that the measurement ranges align with those specified by their manufacturers, with percentage differences with respect to the reference monitor of between 5% and 16%. The diversity found for response time is remarkable, with values ranging from 1 to 15 h, with Pearson correlation factors between 0.63 and 0.90.

Citing Articles

Strategy and Metrological Support for Indoor Radon Measurements Using Popular Low-Cost Active Monitors with High and Low Sensitivity.

Tsapalov A, Kovler K, Bossew P Sensors (Basel). 2024; 24(15).

PMID: 39123814 PMC: 11314910. DOI: 10.3390/s24154764.

References
1.
Wiedner H, Maringer F, Stietka M . Research summary of the EMPIR MetroRADON project. Appl Radiat Isot. 2023; 193:110672. DOI: 10.1016/j.apradiso.2023.110672. View

2.
Rabago D, Quindos L, Vargas A, Sainz C, Radulescu I, Ioan M . Intercomparison of Radon Flux Monitors at Low and at High Radium Content Areas under Field Conditions. Int J Environ Res Public Health. 2022; 19(7). PMC: 8998188. DOI: 10.3390/ijerph19074213. View

3.
Rabago D, Fuente I, Celaya S, Fernandez A, Fernandez E, Quindos J . Intercomparison of Indoor Radon Measurements Under Field Conditions In the Framework of MetroRADON European Project. Int J Environ Res Public Health. 2020; 17(5). PMC: 7084476. DOI: 10.3390/ijerph17051780. View

4.
Tsapalov A, Kovler K . Metrology for Indoor Radon Measurements and Requirements for Different Types of Devices. Sensors (Basel). 2024; 24(2). PMC: 10821384. DOI: 10.3390/s24020504. View

5.
Beck T, Antohe A, Cardellini F, Cucos A, Fialova E, Grossi C . The Metrological Traceability, Performance and Precision of European Radon Calibration Facilities. Int J Environ Res Public Health. 2021; 18(22). PMC: 8623628. DOI: 10.3390/ijerph182212150. View