Functionality Testing of an Innovative Biomechanically Optimized and Surface-modified Orthodontic Mini-screw-a Comparative Study

Overview
Journal J Orofac Orthop
Specialty Dentistry
Date 2024 Jan 15
PMID 38224419
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The failure rate of orthodontic mini-screws depends strongly on primary stability and, thus, on insertion torque. Further improvement regarding the failure rate might be achieved by modifying the surface coating. Therefore, the aim of the study was to investigate the stability of a newly designed and surface-modified orthodontic mini-screw in beagle dogs.

Methods: Newly designed mini-screws coated either with DOTIZE® or DOTIZE®-copper (DOT GmbH, Rostock, Germany; each: n = 24) were inserted in the mandibles of eight beagle dogs for a duration of 8 months. Insertion and removal torque were measured. These data were compared to values generated by using the artificial bone material Sawbones® (Sawbones Europe AB, Malmö, Sweden). Experiments with and without torque limitation (each: n = 5) were run. The bone-to-implant contact rate and the amount of bone between the threads were examined. Statistical significance was set at P < 0.05.

Results: The success rates of the in vivo study reached high levels with 95.3% for the DOTIZE-coated and 90.5% for the DOTIZE-copper-coated screws, whereas the insertion and removal torque did not differ between the coatings. During insertion, a torque limitation of 20 Ncm was necessary to ensure that the recommended limit was not exceeded. The insertion in Sawbones without torque limitation revealed a significantly higher torque compared to torque-limited insertion (18.2 ± 1.3 Ncm, 23.6 ± 1.3 Ncm). Bending occurred (n = 5) in the thread-free part of the mini-screw.

Conclusions: Surface coating might be able to improve the performance of orthodontic mini-screws. The study showed high success rates and stable mini-screws until the end of observation. Further investigations are necessary.

References
1.
Azeem M, Saleem M, Liaquat A, Haq A, Hamid W, Masood M . Failure rates of mini-implants inserted in the retromolar area. Int Orthod. 2019; 17(1):53-59. DOI: 10.1016/j.ortho.2019.01.006. View

2.
Bergemann C, Zaatreh S, Wegner K, Arndt K, Podbielski A, Bader R . Copper as an alternative antimicrobial coating for implants - An study. World J Transplant. 2017; 7(3):193-202. PMC: 5487309. DOI: 10.5500/wjt.v7.i3.193. View

3.
Ten Broeke R, Alves A, Baumann A, Arts J, Geesink R . Bone reaction to a biomimetic third-generation hydroxyapatite coating and new surface treatment for the Symax hip stem. J Bone Joint Surg Br. 2011; 93(6):760-8. DOI: 10.1302/0301-620X.93B6.24986. View

4.
Burghardt I, Luthen F, Prinz C, Kreikemeyer B, Zietz C, Neumann H . A dual function of copper in designing regenerative implants. Biomaterials. 2015; 44:36-44. DOI: 10.1016/j.biomaterials.2014.12.022. View

5.
Chang J, Chen Y, Tung Y, Chiang Y, Lai E, Chen W . Effects of thread depth, taper shape, and taper length on the mechanical properties of mini-implants. Am J Orthod Dentofacial Orthop. 2012; 141(3):279-288. DOI: 10.1016/j.ajodo.2011.09.008. View