Intra-individual Comparison of Coronary CT Angiography-based FFR Between Energy-integrating and Photon-counting Detector CT Systems
Overview
Authors
Affiliations
Background: Coronary computed tomography angiography (CCTA)-based fractional flow reserve (CT-FFR) allows for noninvasive determination of the functional severity of anatomic lesions in patients with coronary artery disease. The aim of this study was to intra-individually compare CT-FFR between photon-counting detector (PCD) and conventional energy-integrating detector (EID) CT systems.
Methods: In this single-center prospective study, subjects who underwent clinically indicated CCTA on an EID-CT system were recruited for a research CCTA on PCD-CT within 30 days. Image reconstruction settings were matched as closely as possible between EID-CT (Bv36 kernel, iterative reconstruction strength level 3, slice thickness 0.5 mm) and PCD-CT (Bv36 kernel, quantum iterative reconstruction level 3, virtual monoenergetic level 55 keV, slice thickness 0.6 mm). CT-FFR was measured semi-automatically using a prototype on-site machine learning algorithm by two readers. CT-FFR analysis was performed per-patient and per-vessel, and a CT-FFR ≤ 0.75 was considered hemodynamically significant.
Results: A total of 22 patients (63.3 ± 9.2 years; 7 women) were included. Median time between EID-CT and PCD-CT was 5.5 days. Comparison of CT-FFR values showed no significant difference and strong agreement between EID-CT and PCD-CT in the per-vessel analysis (0.88 [0.74-0.94] vs. 0.87 [0.76-0.93], P = 0.096, mean bias 0.02, limits of agreement [LoA] -0.14/0.19, r = 0.83, ICC = 0.92), and in the per-patient analysis (0.81 [0.60-0.86] vs. 0.76 [0.64-0.86], P = 0.768, mean bias 0.02, LoA -0.15/0.19, r = 0.90, ICC = 0.93). All included patients were classified into the same category (CT-FFR > 0.75 vs ≤0.75) with both CT systems.
Conclusions: CT-FFR evaluation is feasible with PCD-CT and it shows a strong agreement with EID-CT-based evaluation when images are similarly reconstructed.
Tremamunno G, Pinos D, Zsarnoczay E, Schoepf U, Vecsey-Nagy M, Gnasso C Insights Imaging. 2025; 16(1):36.
PMID: 39961979 PMC: 11832851. DOI: 10.1186/s13244-025-01927-5.
Value of Ultrahigh-Resolution Photon-Counting Detector Computed Tomography in Cardiac Imaging.
Kravchenko D, Hagar M, Vecsey-Nagy M, Tremamunno G, Szilveszter B, Vattay B Echocardiography. 2025; 42(2):e70100.
PMID: 39945196 PMC: 11822751. DOI: 10.1111/echo.70100.
Fahrni G, Boccalini S, Lacombe H, de Oliveira F, Houmeau A, Francart F Eur Radiol. 2024; .
PMID: 39661149 DOI: 10.1007/s00330-024-11237-x.
Photon-counting detector computed tomography in cardiac imaging.
Sharma S, Lemmens M, Smulders M, Budde R, Hirsch A, Mihl C Neth Heart J. 2024; 32(11):405-416.
PMID: 39356451 PMC: 11502613. DOI: 10.1007/s12471-024-01904-5.