» Articles » PMID: 38129900

Biomechanical Analysis of the Door-shaped Titanium Plate in Single-level Anterior Cervical Discectomy and Fusion

Overview
Publisher Biomed Central
Specialty Orthopedics
Date 2023 Dec 22
PMID 38129900
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Analyse and discuss the immediate stability of the cervical spine after anterior cervical discectomy and fusion using a door-shaped titanium plate and compare it with the traditional titanium plate, to provide biomechanical evidence for the rationality and effectiveness of the door-shaped titanium plate in clinical applications.

Methods: Ten adult goat C4/5 vertebral bodies were obtained, and models were prepared using denture base resin. Biomechanical experiments were performed on the specimens before internal fixation. MTS was used to conduct non-destructive biomechanical loading tests in six directions, including flexion, extension, left-right bending, and left-right torsion, recording the range of motion (ROM) and neutral zone (NZ) of each specimen. The specimens were then randomly divided into two groups: the study group was fixed with a door-shaped titanium plate, and the control group was fixed with a traditional titanium plate. ROM and NZ in each direction were measured again. After measurements, both groups were subjected to 0.5 Hz torsion loading with a torque of 2 N m for a total of 3000 cycles, followed by measuring ROM and NZ in six directions once more.

Results: Compared to before fixation, ROM and NZ in both groups significantly decreased in all six directions after fixation, with statistical significance (P < 0.05); after fixation, the study group showed slightly lower values for various mechanical reference parameters compared to the control group, with no statistical significance (P > 0.05); after 3000 torsional loads, both internal fixation groups showed increased ROM and NZ compared to after fixation but to a lower extent, and no screw or titanium plate loosening was observed. Compared to before fixation, the differences were still statistically significant (P < 0.05), with the study group having slightly lower ROM and NZ values in all directions compared to the control group, with no statistical significance (P > 0.05).

Conclusion: The door-shaped titanium plate exhibits mechanical properties similar to the traditional titanium plate in all directions, and its smaller size and simpler surgical operation can be used for anterior cervical endoscopic surgery, reducing surgical trauma. It is clinically feasible and deserves further research and promotion.

Citing Articles

Correlation of transverse rotation of the spine using surface topography and 3D reconstructive radiography in children with idiopathic scoliosis.

Patel M, Liu X, Tassone C, Escott B, Yang K, Thometz J Spine Deform. 2024; 12(4):1001-1008.

PMID: 38403800 DOI: 10.1007/s43390-024-00838-7.

References
1.
Park J, Watthanaaphisit T, Riew K . Timing of development of adjacent-level ossification after anterior cervical arthrodesis with plates. Spine J. 2007; 7(6):633-6. DOI: 10.1016/j.spinee.2006.10.021. View

2.
Butler J, Morrissey P, Wagner S, Kaye I, Sebastian A, Schroeder G . Surgical Strategies to Prevent Adjacent Segment Disease in the Cervical Spine. Clin Spine Surg. 2018; 32(3):91-97. DOI: 10.1097/BSD.0000000000000632. View

3.
Chang B, Feng H, Yu C, Huang K, Gao X, Tang H . [Comparison of clinical effects between anterior cervical Zero-incision fusion system and traditional nail plate system in the treatment of cervical spondylotic myelopathy]. Zhongguo Gu Shang. 2018; 30(5):411-416. DOI: 10.3969/j.issn.1003-0034.2017.05.005. View

4.
Gercek E, Arlet V, Delisle J, Marchesi D . Subsidence of stand-alone cervical cages in anterior interbody fusion: warning. Eur Spine J. 2003; 12(5):513-6. PMC: 3468003. DOI: 10.1007/s00586-003-0539-6. View

5.
Eco L, Brayton A, Whitehead W, Jea A . Reconstruction of the anterior craniocervical junction using an expandable cage after resection of a C1 chordoma in a 5-year-old child: case report. J Neurosurg Pediatr. 2019; 24(1):62-65. DOI: 10.3171/2019.2.PEDS18752. View