Validating a Consumer Smartwatch for Nocturnal Respiratory Rate Measurements in Sleep Monitoring

Overview
Journal Sensors (Basel)
Publisher MDPI
Specialty Biotechnology
Date 2023 Sep 28
PMID 37766031
Authors
Affiliations
Soon will be listed here.
Abstract

Wrist-based respiratory rate (RR) measurement during sleep faces accuracy limitations. This study aimed to assess the accuracy of the RR estimation function during sleep based on the severity of obstructive sleep apnea (OSA) using the Samsung Galaxy Watch (GW) series. These watches are equipped with accelerometers and photoplethysmography sensors for RR estimation. A total of 195 participants visiting our sleep clinic underwent overnight polysomnography while wearing the GW, and the RR estimated by the GW was compared with the reference RR obtained from the nasal thermocouple. For all participants, the root mean squared error (RMSE) of the average overnight RR and continuous RR measurements were 1.13 bpm and 1.62 bpm, respectively, showing a small bias of 0.39 bpm and 0.37 bpm, respectively. The Bland-Altman plots indicated good agreement in the RR measurements for the normal, mild, and moderate OSA groups. In participants with normal-to-moderate OSA, both average overnight RR and continuous RR measurements achieved accuracy rates exceeding 90%. However, for patients with severe OSA, these accuracy rates decreased to 79.45% and 75.8%, respectively. The study demonstrates the GW's ability to accurately estimate RR during sleep, even though accuracy may be compromised in patients with severe OSA.

References
1.
Dehkordi P, Garde A, Molavi B, Ansermino J, Dumont G . Extracting Instantaneous Respiratory Rate From Multiple Photoplethysmogram Respiratory-Induced Variations. Front Physiol. 2018; 9:948. PMC: 6058306. DOI: 10.3389/fphys.2018.00948. View

2.
Karlen W, Raman S, Ansermino J, Dumont G . Multiparameter respiratory rate estimation from the photoplethysmogram. IEEE Trans Biomed Eng. 2013; 60(7):1946-53. DOI: 10.1109/TBME.2013.2246160. View

3.
Do W, Russell R, Wheeler C, Lockwood M, De Vos M, Pavord I . Performance of Contactless Respiratory Rate Monitoring by Albus Home, an Automated System for Nocturnal Monitoring at Home: A Validation Study. Sensors (Basel). 2022; 22(19). PMC: 9573065. DOI: 10.3390/s22197142. View

4.
Knorr-Chung B, McGrath S, Blike G . Identifying airway obstructions using photoplethysmography (PPG). J Clin Monit Comput. 2008; 22(2):95-101. DOI: 10.1007/s10877-008-9110-7. View

5.
Ballal T, Heneghan C, Zaffaroni A, Boyle P, de Chazal P, Shouldice R . A pilot study of the nocturnal respiration rates in COPD patients in the home environment using a non-contact biomotion sensor. Physiol Meas. 2014; 35(12):2513-27. DOI: 10.1088/0967-3334/35/12/2513. View