Three-dimensional Echo-shifted EPI with Simultaneous Blip-up and Blip-down Acquisitions for Correcting Geometric Distortion

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2023 Sep 5
PMID 37667533
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: EPI with blip-up/down acquisition (BUDA) can provide high-quality images with minimal distortions by using two readout trains with opposing phase-encoding gradients. Because of the need for two separate acquisitions, BUDA doubles the scan time and degrades the temporal resolution when compared to single-shot EPI, presenting a major challenge for many applications, particularly fMRI. This study aims at overcoming this challenge by developing an echo-shifted EPI BUDA (esEPI-BUDA) technique to acquire both blip-up and blip-down datasets in a single shot.

Methods: A 3D esEPI-BUDA pulse sequence was designed by using an echo-shifting strategy to produce two EPI readout trains. These readout trains produced a pair of k-space datasets whose k-space trajectories were interleaved with opposite phase-encoding gradient directions. The two k-space datasets were separately reconstructed using a 3D SENSE algorithm, from which time-resolved B -field maps were derived using TOPUP in FSL and then input into a forward model of joint parallel imaging reconstruction to correct for geometric distortion. In addition, Hankel structured low-rank constraint was incorporated into the reconstruction framework to improve image quality by mitigating the phase errors between the two interleaved k-space datasets.

Results: The 3D esEPI-BUDA technique was demonstrated in a phantom and an fMRI study on healthy human subjects. Geometric distortions were effectively corrected in both phantom and human brain images. In the fMRI study, the visual activation volumes and their BOLD responses were comparable to those from conventional 3D echo-planar images.

Conclusion: The improved imaging efficiency and dynamic distortion correction capability afforded by 3D esEPI-BUDA are expected to benefit many EPI applications.

References
1.
Griswold M, Jakob P, Heidemann R, Nittka M, Jellus V, Wang J . Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002; 47(6):1202-10. DOI: 10.1002/mrm.10171. View

2.
Lin F, Huang T, Chen N, Wang F, Stufflebeam S, Belliveau J . Functional MRI using regularized parallel imaging acquisition. Magn Reson Med. 2005; 54(2):343-53. PMC: 3779863. DOI: 10.1002/mrm.20555. View

3.
Pruessmann K, Weiger M, Scheidegger M, Boesiger P . SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999; 42(5):952-62. View

4.
Andersson J, Skare S, Ashburner J . How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging. Neuroimage. 2003; 20(2):870-88. DOI: 10.1016/S1053-8119(03)00336-7. View

5.
In M, Posnansky O, Beall E, Lowe M, Speck O . Distortion correction in EPI using an extended PSF method with a reversed phase gradient approach. PLoS One. 2015; 10(2):e0116320. PMC: 4338274. DOI: 10.1371/journal.pone.0116320. View