Intra-individual Comparison of Image Quality of the Coronary Arteries Between Photon-counting Detector and Energy-integrating Detector CT Systems
Overview
Authors
Affiliations
Purpose: To intra-individually compare the objective and subjective image quality of coronary computed tomography angiography (CCTA) between photon-counting detector CT (PCD-CT) and energy-integrating detector CT (EID-CT).
Method: Consecutive patients undergoing clinically indicated CCTA on an EID-CT system were prospectively enrolled for a research CCTA performed on a PCD-CT system within 30 days. Polychromatic images were reconstructed for both EID- and PCD-CT, while virtual monoenergetic images (VMI) were generated at 40, 45, 50, 55, 60 and 70 keV for PCD-CT. Two blinded readers calculated contrast-to-noise ratio (CNR) for each major coronary artery and rated image noise, vessel attenuation, vessel sharpness, and overall quality on a 1-5 Likert scale. Patients were then stratified by body mass index (BMI) [high (>30 kg/m) vs low (<30 kg/m)] for subgroup analysis.
Results: A total of 20 patients (67.5 ± 9.0 years, 75% male) were included in the study. Compared with EID-CT, coronary artery CNR values from PCD-CT monoenergetic and polychromatic reconstructions were all significantly higher than CNR values from EID-CT, with incrementally greater differences in obese subjects (all p < 0.008). Subjective image noise and sharpness were also significantly higher for all VMI reconstructions compared to EID-CT (all p < 0.008). All subjective scores were significantly higher for 55, 60, and 70 keV PCD-CT than EID-CT values (all p < 0.05).
Conclusions: The improved objective and subjective image quality of PCD-CT compared to EID-CT may provide better visualization of the coronary arteries for a wide array of patients, especially those with a high BMI.
Tremamunno G, Pinos D, Zsarnoczay E, Schoepf U, Vecsey-Nagy M, Gnasso C Insights Imaging. 2025; 16(1):36.
PMID: 39961979 PMC: 11832851. DOI: 10.1186/s13244-025-01927-5.
Multienergy cardiovascular CT imaging: current state and future.
Klambauer K, Lisi C, Moser L, Mergen V, Flohr T, Eberhard M Br J Radiol. 2024; 98(1167):321-329.
PMID: 39656967 PMC: 11840172. DOI: 10.1093/bjr/tqae246.
Caruso D, De Santis D, Tremamunno G, Santangeli C, Polidori T, Bona G Eur Radiol. 2024; .
PMID: 39299952 DOI: 10.1007/s00330-024-11059-x.
Meloni A, Maffei E, Positano V, Clemente A, De Gori C, Berti S Cardiovasc Diagn Ther. 2024; 14(4):698-724.
PMID: 39263472 PMC: 11384460. DOI: 10.21037/cdt-24-52.
Lacaita P, Luger A, Troger F, Widmann G, Feuchtner G J Cardiovasc Dev Dis. 2024; 11(4).
PMID: 38667745 PMC: 11050624. DOI: 10.3390/jcdd11040127.