» Articles » PMID: 37317675

Bloch Simulator-driven Deep Recurrent Neural Network for Magnetization Transfer Contrast MR Fingerprinting and CEST Imaging

Overview
Journal Magn Reson Med
Publisher Wiley
Specialty Radiology
Date 2023 Jun 15
PMID 37317675
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To develop a unified deep-learning framework by combining an ultrafast Bloch simulator and a semisolid macromolecular magnetization transfer contrast (MTC) MR fingerprinting (MRF) reconstruction for estimation of MTC effects.

Methods: The Bloch simulator and MRF reconstruction architectures were designed with recurrent neural networks and convolutional neural networks, evaluated with numerical phantoms with known ground truths and cross-linked bovine serum albumin phantoms, and demonstrated in the brain of healthy volunteers at 3 T. In addition, the inherent magnetization-transfer ratio asymmetry effect was evaluated in MTC-MRF, CEST, and relayed nuclear Overhauser enhancement imaging. A test-retest study was performed to evaluate the repeatability of MTC parameters, CEST, and relayed nuclear Overhauser enhancement signals estimated by the unified deep-learning framework.

Results: Compared with a conventional Bloch simulation, the deep Bloch simulator for generation of the MTC-MRF dictionary or a training data set reduced the computation time by 181-fold, without compromising MRF profile accuracy. The recurrent neural network-based MRF reconstruction outperformed existing methods in terms of reconstruction accuracy and noise robustness. Using the proposed MTC-MRF framework for tissue-parameter quantification, the test-retest study showed a high degree of repeatability in which the coefficients of variance were less than 7% for all tissue parameters.

Conclusion: Bloch simulator-driven, deep-learning MTC-MRF can provide robust and repeatable multiple-tissue parameter quantification in a clinically feasible scan time on a 3T scanner.

Citing Articles

Improving quantification accuracy of a nuclear Overhauser enhancement signal at -1.6 ppm at 4.7 T using a machine learning approach.

Yin L, Viswanathan M, Kurmi Y, Zu Z Phys Med Biol. 2025; 70(2.

PMID: 39774035 PMC: 11740009. DOI: 10.1088/1361-6560/ada716.


Synthesis of MR fingerprinting information from magnitude-only MR imaging data using a parallelized, multi network U-Net convolutional neural network.

McGee K, Sui Y, Witte R, Panda A, Campeau N, Mostardeiro T Front Radiol. 2025; 4:1498411.

PMID: 39742349 PMC: 11686891. DOI: 10.3389/fradi.2024.1498411.


mapping of the chemical exchange relayed nuclear Overhauser effect using deep magnetic resonance fingerprinting.

Power I, Rivlin M, Shmuely H, Zaiss M, Navon G, Perlman O iScience. 2024; 27(11):111209.

PMID: 39569380 PMC: 11576397. DOI: 10.1016/j.isci.2024.111209.


Enhancing amide proton transfer imaging in ischemic stroke using a machine learning approach with partially synthetic data.

Viswanathan M, Yin L, Kurmi Y, Afzal A, Zu Z NMR Biomed. 2024; 38(1):e5277.

PMID: 39434444 PMC: 11602689. DOI: 10.1002/nbm.5277.


Unraveling contributions to the Z-spectrum signal at 3.5 ppm of human brain tumors.

Heo H, Singh M, Mahmud S, Blair L, Kamson D, Zhou J Magn Reson Med. 2024; 92(6):2641-2651.

PMID: 39086185 PMC: 11436306. DOI: 10.1002/mrm.30241.


References
1.
Quesson B, Thiaudiere E, Delalande C, Chateil J, Moonen C, Canioni P . Magnetization transfer imaging of rat brain under non-steady-state conditions. Contrast prediction using a binary spin-bath model and a super-lorentzian lineshape. J Magn Reson. 1998; 130(2):321-8. DOI: 10.1006/jmre.1997.1326. View

2.
Morrison C, Henkelman R . A model for magnetization transfer in tissues. Magn Reson Med. 1995; 33(4):475-82. DOI: 10.1002/mrm.1910330404. View

3.
Zhou Y, Bie C, van Zijl P, Yadav N . The relayed nuclear Overhauser effect in magnetization transfer and chemical exchange saturation transfer MRI. NMR Biomed. 2022; 36(6):e4778. PMC: 9708952. DOI: 10.1002/nbm.4778. View

4.
Heo H, Zhang Y, Lee D, Hong X, Zhou J . Quantitative assessment of amide proton transfer (APT) and nuclear overhauser enhancement (NOE) imaging with extrapolated semi-solid magnetization transfer reference (EMR) signals: Application to a rat glioma model at 4.7 Tesla. Magn Reson Med. 2015; 75(1):137-49. PMC: 4561043. DOI: 10.1002/mrm.25581. View

5.
Wen Q, Wang K, Hsu Y, Xu Y, Sun Y, Wu D . Chemical exchange saturation transfer imaging for epilepsy secondary to tuberous sclerosis complex at 3 T: Optimization and analysis. NMR Biomed. 2021; 34(9):e4563. DOI: 10.1002/nbm.4563. View