How and when to Measure PH in IVF Culture Media: Validation of a Portable Blood Gas Analyzer in Two IVF Culture Dishes for Time Lapse and Conventional Incubators

Overview
Publisher Springer
Date 2023 Jun 14
PMID 37314570
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Maintaining a stable pH at optimal level in human embryo culture media is crucial for embryo development but poses a challenge for all IVF laboratories. We validate analytically reliable conditions for pH measurement that are as close as possible to the embryo microenvironment during IVF.

Methods: This was a multicentric study. A Siemens EPOC portable blood gas analyzer was used. The analytical validation was carried out under the culture medium (Global Total HSA®) conditions of use (microdroplets, under oil overlay, in a IVF incubator with (EmbryoScope®) or without a time lapse system (K system G210+®) and using IVF dishes. The validation included repeatability ("within-run" precision), total precision (between-day precision), trueness based on inter-laboratory comparison, inaccuracy based on external quality assessment and comparison to the reference technique. We also assessed the pre-analytical medium incubation time required to obtain a target value.

Results: A measurement after an incubation period of 24 to 48 h is more representative of the pH to which the embryo will be exposed throughout the culture. The "within-run" and "between-day" precision show very low coefficients of variation (CV%): 0.17 to 0.22% and 0.13 to 0.34%, respectively, with IVF culture media. Trueness (% bias) range from - 0.07 to - 0.03%. We demonstrate good correlation between EPOC and reference pH electrode with an overestimation of 0.03 pH units of EPOC.

Conclusion: Our method demonstrates good analytical performance for IVF laboratories wishing to implement a robust quality assurance system to monitor pH in embryo culture media. Compliance with stringent pre-analytical and analytical conditions is essential.

Citing Articles

Towards homogenous multiwell plate based pH sensors using a responsive triangulenium dye and an ATTO-647 reference dye.

Wied M, Sorensen T RSC Adv. 2025; 15(2):1438-1446.

PMID: 39822565 PMC: 11736390. DOI: 10.1039/d4ra07949c.

References
1.
Squirrell J, Lane M, Bavister B . Altering intracellular pH disrupts development and cellular organization in preimplantation hamster embryos. Biol Reprod. 2001; 64(6):1845-54. PMC: 5087321. DOI: 10.1095/biolreprod64.6.1845. View

2.
Luukkonen A, Lehto T, Hedberg P, Vaskivuo T . Evaluation of a hand-held blood gas analyzer for rapid determination of blood gases, electrolytes and metabolites in intensive care setting. Clin Chem Lab Med. 2015; 54(4):585-94. DOI: 10.1515/cclm-2015-0592. View

3.
Gatimel N, Moreau J, Parinaud J, Leandri R . Need for choosing the ideal pH value for IVF culture media. J Assist Reprod Genet. 2020; 37(5):1019-1028. PMC: 7244657. DOI: 10.1007/s10815-020-01726-5. View

4.
Edwards L, Williams D, Gardner D . Intracellular pH of the preimplantation mouse embryo: effects of extracellular pH and weak acids. Mol Reprod Dev. 1998; 50(4):434-42. DOI: 10.1002/(SICI)1098-2795(199808)50:4<434::AID-MRD7>3.0.CO;2-J. View

5.
Edwards L, Williams D, Gardner D . Intracellular pH of the mouse preimplantation embryo: amino acids act as buffers of intracellular pH. Hum Reprod. 1999; 13(12):3441-8. DOI: 10.1093/humrep/13.12.3441. View