4D-Flow Cardiovascular Magnetic Resonance Sequence for Aortic Assessment: Multi-Vendor and Multi-Magnetic Field Reproducibility in Healthy Volunteers

Overview
Journal J Clin Med
Specialty General Medicine
Date 2023 Apr 28
PMID 37109295
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: Four-dimensional (4D) flow cardiac magnetic resonance (CMR) represents an emerging technique for non-invasive evaluation of the aortic flow. The aim of this study was to investigate a 4D-flow CMR sequence for the assessment of thoracic aorta comparing different vendors and different magnetic fields of MR scanner in fifteen healthy volunteers.

Methods: CMR was performed on three different MRI scanners: one at 1.5 T and two at 3 T. Flow parameters and planar wall shear stress (WSS) were extracted from six transversal planes along the full thoracic aorta by three operators. Inter-vendor comparability as well as scan-rescan, intra- and interobserver reproducibility were examined.

Results: A high heterogeneity was found in the comparisons for each operator and for each scanner in the six transversal planes analysis (Friedman rank-sum test; -value ≤ 0.05). Among all, the most reproducible measures were extracted for the sinotubular junction plane and for the flow parameters.

Conclusions: Our results suggest that standardized procedures have to be defined to make more comparable and reproducible 4D-flow parameters and mainly, clinical impactfulness. Further studies on sequences development are needed to validate 4D-flow MRI assessment across vendors and magnetic fields also compared to a missing gold standard.

Citing Articles

Inter-site comparability of 4D flow cardiovascular magnetic resonance measurements in healthy traveling volunteers-a multi-site and multi-magnetic field strength study.

Muller M, Daud E, Langer G, Groschel J, Viezzer D, Hadler T Front Cardiovasc Med. 2024; 11:1456814.

PMID: 39582524 PMC: 11582008. DOI: 10.3389/fcvm.2024.1456814.

References
1.
Hope M, Hope T, Meadows A, Ordovas K, Urbania T, Alley M . Bicuspid aortic valve: four-dimensional MR evaluation of ascending aortic systolic flow patterns. Radiology. 2010; 255(1):53-61. DOI: 10.1148/radiol.09091437. View

2.
Geeraert P, Jamalidinan F, Hassanabad A, Sojoudi A, Bristow M, Lydell C . Bicuspid aortic valve disease is associated with abnormal wall shear stress, viscous energy loss, and pressure drop within the ascending thoracic aorta: A cross-sectional study. Medicine (Baltimore). 2021; 100(26):e26518. PMC: 8257908. DOI: 10.1097/MD.0000000000026518. View

3.
Ebel S, Hubner L, Kohler B, Kropf S, Preim B, Jung B . Validation of two accelerated 4D flow MRI sequences at 3 T: a phantom study. Eur Radiol Exp. 2019; 3(1):10. PMC: 6391502. DOI: 10.1186/s41747-019-0089-2. View

4.
Rodriguez-Palomares J, Dux-Santoy L, Guala A, Kale R, Maldonado G, Teixido-Tura G . Aortic flow patterns and wall shear stress maps by 4D-flow cardiovascular magnetic resonance in the assessment of aortic dilatation in bicuspid aortic valve disease. J Cardiovasc Magn Reson. 2018; 20(1):28. PMC: 5918697. DOI: 10.1186/s12968-018-0451-1. View

5.
Farag E, van Ooij P, Planken R, Dukker K, de Heer F, Bouma B . Aortic valve stenosis and aortic diameters determine the extent of increased wall shear stress in bicuspid aortic valve disease. J Magn Reson Imaging. 2018; 48(2):522-530. PMC: 6099246. DOI: 10.1002/jmri.25956. View