Noninvasive Continuous Glucose Monitoring With a Novel Wearable Dial Resonating Sensor: A Clinical Proof-of-Concept Study

Overview
Specialty Endocrinology
Date 2023 Apr 27
PMID 37102600
Authors
Affiliations
Soon will be listed here.
Abstract

Background: A noninvasive, wearable continuous glucose monitor would be a major advancement in diabetes therapy. This trial investigated a novel noninvasive glucose monitor which analyzes spectral variations in radio frequency/microwave signals reflected from the wrist.

Methods: A single-arm, open-label, experimental study compared glucose values from a prototype investigational device with laboratory glucose measurements from venous blood samples (Super GL Glucose Analyzer, Dr. Müller Gerätebau GmbH) at varying levels of glycemia. The study included 29 male participants with type 1 diabetes (age range = 19-56 years). The study comprised three stages with the following aims: (1) demonstrate initial proof-of-principle, (2) test an improved device design, and (3) test performance on two consecutive days without device recalibration. The co-primary endpoints in all trial stages were median and mean absolute relative difference (ARD) calculated across all data points.

Results: In stage 1, the median and mean ARDs were 30% and 46%, respectively. Stage 2 produced marked performance improvements with a median and mean ARD of 22% and 28%, respectively. Stage 3 showed that, without recalibration, the device performed as well as the initial prototype (stage 1) with a median and mean ARD of 35% and 44%, respectively.

Conclusion: This proof-of-concept study shows that a novel noninvasive continuous glucose monitor was capable of detecting glucose levels. Furthermore, the ARD results are comparable to first models of commercially available minimally invasive products without the need to insert a needle. The prototype has been further developed and is being tested in subsequent studies.

Trial Registration Number: NCT05023798.

Citing Articles

Using Artificial Intelligence to Improve the Accuracy of a Wrist-Worn, Noninvasive Glucose Monitor: A Pilot Study.

Qureshi M, Bain S, Luzio S, Handy C, Fowles D, Love B J Diabetes Sci Technol. 2024; :19322968241252819.

PMID: 38757895 PMC: 11571554. DOI: 10.1177/19322968241252819.


Blood glucose monitoring devices for type 1 diabetes: a journey from the food and drug administration approval to market availability.

Mittal R, Koutras N, Maya J, Lemos J, Hirani K Front Endocrinol (Lausanne). 2024; 15:1352302.

PMID: 38559693 PMC: 10978642. DOI: 10.3389/fendo.2024.1352302.


Beware of buying online noninvasive continuous glucose monitoring devices: danger is just around the corner!.

Gamarra E, Barchi G, Camponovo C, Trimboli P Acta Diabetol. 2023; 61(2):253-255.

PMID: 37776376 DOI: 10.1007/s00592-023-02189-7.

References
1.
Heinemann L, Krisiunas E . Diabetes Technology and Waste: A Complex Problem Piling Up!. J Diabetes Sci Technol. 2019; 13(5):815-816. PMC: 6955464. DOI: 10.1177/1932296819836395. View

2.
Floyd B, Chandra P, Hall S, Phillips C, Alema-Mensah E, Strayhorn G . Comparative analysis of the efficacy of continuous glucose monitoring and self-monitoring of blood glucose in type 1 diabetes mellitus. J Diabetes Sci Technol. 2012; 6(5):1094-102. PMC: 3570843. DOI: 10.1177/193229681200600513. View

3.
Hoss U, Budiman E, Liu H, Christiansen M . Continuous glucose monitoring in the subcutaneous tissue over a 14-day sensor wear period. J Diabetes Sci Technol. 2013; 7(5):1210-9. PMC: 3876365. DOI: 10.1177/193229681300700511. View

4.
Hoss U, Budiman E, Liu H, Christiansen M . Feasibility of Factory Calibration for Subcutaneous Glucose Sensors in Subjects With Diabetes. J Diabetes Sci Technol. 2014; 8(1):89-94. PMC: 4454101. DOI: 10.1177/1932296813511747. View

5.
Messer L, Berget C, Beatson C, Polsky S, Forlenza G . Preserving Skin Integrity with Chronic Device Use in Diabetes. Diabetes Technol Ther. 2018; 20(S2):S254-S264. PMC: 6011799. DOI: 10.1089/dia.2018.0080. View